临床验证M-TONX:一种新的复合反弹音标仪和心智仪
Naveed Nilforushan1, Mozafar Yousefi2, Ankica Babic3,4
1Eye Research Centre, Iran University of Medical Sciences, Tehran, Iran.
Translational vision science & technology
|December 26, 2024
概括
M-TONX设备可同时可靠地测量眼内压力 (IOP) 和中角膜厚度 (CCT). 它的准确性在广泛的测量中保持稳定,证实了其临床实用性.
科学领域:
- 眼科医生 眼科 眼科
- 生物医学工程 生物医学工程
- 医疗器械 医疗器械
背景情况:
- 精确测量眼内压力 (IOP) 和中角膜厚度 (CCT) 对于诊断和管理青光眼和其他眼部疾病至关重要.
- 目前的方法通常需要单独的设备,增加检查时间和复杂性.
- 该M-TONX设备使用创新的反弹技术同时进行IOP和CCT测量.
研究的目的:
- 临床研究M-TONX设备的性能,用于同时测量IOP和CCT.
- 为了将M-TONX测量与既定标准设备进行比较,例如高盛平面化音量计 (GAT) 和Pentacam角膜拓仪.
- 评估M-TONX在不同IOP和CCT范围内的准确性和可靠性.
主要方法:
- 使用M-TONX设备测量了225名受试者的374只眼睛的内视力和CCT.
- 用高盛平面化度仪 (GAT) 和Pentacam角膜拓仪作为黄金标准重复测量.
- 统计分析涉及基于IOP (<=16,16-23,>=23毫米) 和CCT (<=475,475-574,>=575微米) 的患者分层.
主要成果:
- M-TONX与标准设备具有很高的一致性,IOP的置信区间为0.7-1.2mm Hg,CCT的置信区间为-12.7至-5.4μm.
- 分层分析显示了实质性的一致性,所有组和类的类内相关系数>0.65和皮尔森相关系数>0.8.
- 该设备在广泛的IOP和CCT值中被证明是准确的.
结论:
- 该M-TONX设备表现出可靠的性能,与IOP和CCT测量的标准方法相美.
- 它的准确性在各种IOP和CCT范围内是一致的,表明临床有效性.
- M-TONX提供了一个用户友好的,紧的解决方案,将两个基本的眼科测量集成到一个单一的设备中.
相关概念视频
Assessment of blood pressure in brachial artery(two-step method)
1.3K
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
1.3K
Measurement of Blood Pressure
2.4K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
2.4K
Assessment of blood pressure in brachial artery(one-step method)
988
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
988
Equipments Used To Measure Blood Pressure
2.9K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.9K