建立和应用基于ICP-MS的双免疫测试方法,用于标记稳定元素抗体
Xuyao Ji1, Jing Wu1, Jing Zhang1
1Department of Laboratory Medicine at the Fourth Affiliated Hospital, Harbin Medical University, Harbin, China.
Journal of clinical laboratory analysis
|December 31, 2024
概括
使用诱导合血质谱法 (ICP-MS) 的新型双免疫试验可以同时检测α-fetoprotein (AFP) 和前列腺特异性抗原 (PSA). 这种经过验证的方法显示了临床血清样本的高精度和特异性.
科学领域:
- 分析化学 分析化学
- 生物化学 生化学
- 临床诊断 临床诊断 临床诊断
背景情况:
- 开发一种双重免疫试验,同时检测血清中的α-fetoprotein (AFP) 和前列腺特异性抗原 (PSA).
- 使用感应合等离子体质谱法 (ICP-MS) 与稳定元素标记抗体.
研究的目的:
- 建立和评估一种新的AFP和PSA双重免疫试验.
- 为了验证基于ICP-MS的方法的性能和临床适用性.
主要方法:
- 通过使用磁珠和稀土元素标签 (Europium和Samarium) 建立了双抗体三明治免疫试验.
- 优化了测试条件,并根据CLSI指南进行了性能验证.
- 使用临床血清样本验证了该方法.
主要成果:
- 达到了AFP (0.76 ng/mL) 和PSA (0.41 ng/mL) 的低检测极限 (LOD).
- 在两种分析物中都表现出优异的批量内 (<3.33%) 和批量间 (<4.97%) 精度.
- 与CLIA方法相比,报告了较高的相关系数 (r=0.9930对于AFP,r=0.9852对于PSA).
结论:
- 成功开发了一种基于ICP-MS的双免疫试验,用于同时检测血清中的AFP和PSA.
- 该方法具有良好的精度,宽线性范围和高特异性,满足临床要求.
- 开发的试验显示出在诊断中临床应用的巨大潜力.
相关概念视频
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
630
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
630
Immunogold Electron Microscopy
4.0K
Immunoelectron microscopy utilizes immunogold labeling of endogenous proteins with specific antibodies to detect and localize these proteins in cells and tissues. The procedure provides insights into the distribution and quantification of protein under different stimulation conditions offering clues about their functions. Conjugating highly electron-dense gold particles with primary or secondary antibodies allow antigen detection on and within cells, with high resolution and specificity.
4.0K
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
382
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
382
Atomic Emission Spectroscopy: Lab
148
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
148