电子温度测量从中性原子排放线比率的电子温度测量
Steven F Adams1, Chase S Murray2, Neil A Pohl2
1Air Force Research Laboratory, Wright-Patterson AFB, Ohio 45433, USA.
The Review of scientific instruments
|January 6, 2025
概括
这项研究引入了一种使用原子排放线测量等离子电子温度的新方法. 该技术对激光森散射进行校准,使未来的测量能够在没有这种复杂的设备的情况下进行.
科学领域:
- 等离子体物理学的物理学
- 原子光谱学 原子光谱学
背景情况:
- 精确测量电子温度对于理解等离子体行为至关重要.
- 像激光森散射这样的现有方法可能是复杂和昂贵的.
研究的目的:
- 开发和验证一种用于确定等离子电子温度的新方法.
- 为了利用原子的辐射光谱进行温度诊断.
主要方法:
- 原子排放线 (400.88和401.52 nm) 的光谱分析.
- 测量脉冲空气等离子体在放电后区域的线路强度比.
- 使用激光森散射对的排放方法进行校准.
主要成果:
- 确定了特定线和电子温度的强度比率之间的相关性.
- 导出了一个校准因子,表示爱因斯坦A系数的新比例.
- 的排放方法得到了对激光森散射的验证.
结论:
- 原子发射提供了可行的电子温度在等离子体的诊断.
- 开发的方法为未来的测量提供了一个更简单的替代方案,而不是激光森散射.
- 由此衍生的校准因子使得使用线比率直接确定电子温度.
相关概念视频
Atomic Spectroscopy: Effects of Temperature
226
Atomization, converting samples into gas-phase atoms and ions, is essential for atomic spectroscopy. The flame temperature required for atomization affects the efficiency of the atomic spectroscopic methods by increasing the atomization efficiency and the relative population of the excited and ground states.
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
226
Emission Spectra
48.9K
When solids, liquids, or condensed gases are heated sufficiently, they radiate some of the excess energy as light. Photons produced in this manner have a range of energies, and thereby produce a continuous spectrum in which an unbroken series of wavelengths is present.
48.9K
Atomic Emission Spectroscopy: Overview
508
Atomic emission spectroscopy (AES) is an analytical technique used to determine the elemental composition of a sample by analyzing the light emitted from excited atoms. In AES, atoms in a sample are excited to higher energy levels by thermal energy from high-temperature sources, such as plasma, arcs, or sparks. When these excited atoms return to lower energy states, they emit light at specific wavelengths characteristic of each element. The resulting atomic emission spectrum, which consists of...
508
Atomic Emission Spectroscopy: Lab
122
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
122
Atomic Emission Spectroscopy: Instrumentation
252
The instrumentation of atomic emission spectrometry (AES) involves various components, including atomization devices that convert samples into gas-phase atoms and ions. There are two main types of atomization devices: continuous and discrete atomizers. Continuous atomizers, like plasmas and flames, introduce samples in a constant stream, while discrete atomizers inject individual samples using syringes or autosamplers. The most common discrete atomizer is the electrothermal atomizer.
252
Atomic Nuclei: Nuclear Spin State Population Distribution
885
Near absolute zero temperatures, in the presence of a magnetic field, the majority of nuclei prefer the lower energy spin-up state to the higher energy spin-down state. As temperatures increase, the energy from thermal collisions distributes the spins more equally between the two states. The Boltzmann distribution equation gives the ratio of the number of spins predicted in the spin −½ (N−) and spin +½ (N+) states.
885