验证一个actigraphy手表的睡眠测量:仪器验证研究研究
Mari Waki1, Ryohei Nakada1, Kayo Waki1
1Department of Biomedical Informatics, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan.
JMIR formative research
|January 6, 2025
概括
最初,iAide2设备对睡眠时间和总睡眠时间 (TST) 显示了中等的准确性. 校准显著改善了iAide2 的性能.
科学领域:
- 生物医学工程 生物医学工程
- 睡眠科学 睡眠科学
- 可穿戴技术可穿戴技术
背景情况:
- iAide2 (Tokai) 是一款具有潜在研究应用的无线体力活动监测器.
- 它的睡眠测量准确性与验证的标准相比仍然未被评估.
研究的目的:
- 评估iAide2在测量睡眠时间和总睡眠时间 (TST) 的性能.
- 开发和验证一个校准方法,以提高iAide2的睡眠测量准确度.
主要方法:
- 在6名参与者身上进行了自由生活睡眠监测,使用iAide2和经过验证的动图仪 (MotionWatch 8).
- 睡眠时间和TST在64个晚上被记录下来,两种设备都使用事件标记器同步.
- 类内相关系数 (ICC) 用于在校准前后评估设备之间的一致性.
主要成果:
- 未校准的iAide2显示了对睡眠时间 (ICC=0.69) 和TST (ICC=0.56) 的中度一致.
- 一致性ICCs对于睡眠时间 (0.91) 是很好的,对于TST (0.78) 是中度的.
- 在校准后,ICC显著改善,为睡眠时间的0.96和TST的0.82,在参与者中显示出强度.
结论:
- 未校准的iAide2测量睡眠时间和TST有显著的错误,限制了它们的绝对量化使用.
- 一种简单,强大的校准方法大大提高了iAide2的准确性,验证了其在睡眠研究中的使用.
- 校准的iAide2适用于体力活动研究中的可靠睡眠测量.
相关概念视频
Holter Monitor: 24-Hour Monitoring
1
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
1
Special considerations while measuring blood pressure
703
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
703