消费者可穿戴设备用于检测心房动的比较评估:验证研究
Femke Wouters1,2, Henri Gruwez1,2,3,4, Christophe Smeets1,2
1Limburg Clinical Research Center/Mobile Health Unit, Faculty of Medicine and Life Sciences, Hasselt University, Hasselt, Belgium.
JMIR formative research
|January 10, 2025
概括
消费者可穿戴设备可以准确检测心房动 (AF). 这项研究发现,心电图和PPG设备之间没有显著的性能差异,突出了它们在广泛的AF查方面的潜力.
科学领域:
- 心脏病学 心脏病学
- 生物医学工程 生物医学工程
- 数字健康数字健康
背景情况:
- 智能手机和智能手表等面向消费者的可穿戴设备 (CWD) 越来越多地用于检测心房动 (AF).
- 现有的验证研究众多,但缺乏直接比较设备间性能.
- 这些设备使用心电图 (ECG) 或光电显微镜 (PPG) 进行数字记录.
研究的目的:
- 评估和比较CWD患者在区分鼻腔节律和AF的诊断表现.
- 在临床环境中评估不同CWD的准确性.
主要方法:
- 招募了122名鼻节律或AF的参与者.
- 用KardiaMobile 6L (手持式心电图),Apple Watch (智能手表心电图),FibriCheck (智能手机PPG) 和Preventicus (智能手机PPG) 来进行测量.
- 一个12导心电图作为黄金标准,以随机顺序进行测量.
主要成果:
- 所有测试的设备在检测AF时都显示出100%的灵敏度.
- 检测鼻节律的特异性很高,在各种设备上都是可比的 (KardiaMobile 6L: 96.4%,Apple Watch: 97.8%,FibriCheck: 98.9%,Preventicus: 97.8%).
- 在设备之间没有观察到性能 (P=.50) 或测量质量 (P=.21) 的显著差异,尽管参与者更喜欢Apple Watch.
结论:
- 面向消费者的可穿戴设备在区分鼻腔节律和AF之间表现出很高的准确性.
- 在评估的基于心电图和PPG的CWD中没有发现显著的性能差异.
- 这些发现支持CWD在广泛人群中用于AF检测的实用性.
相关概念视频
Pulse rhythm
758
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
758
Equipments Used To Measure Blood Pressure
821
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
821