在可穿戴技术中验证有氧容量 (VO2max) 和脉冲氧度
Bryson Carrier1, Sofia Marten Chaves1, James W Navalta1
1School of Integrated Health Sciences, Department of Kinesiology and Nutrition Sciences, University of Nevada, Las Vegas, NV 89154, USA.
嘉明fēnix 6准确地估计了VO2max,但其血液氧和 (BOS) 测量值是不可靠的. 这种可穿戴技术对于健身跟踪是有用的,但不是用于医疗监测.
科学领域:
- 运动生理学 运动生理学
- 生物医学工程 生物医学工程
- 可穿戴技术验证的验证
背景情况:
- 可穿戴技术越来越受欢迎,需要独立的准确性验证.
- 评估消费者级生理监测设备的可靠性至关重要.
研究的目的:
- 评估来自Garmin fēnix 6.的VO2max和血液氧和 (BOS) 测量结果的准确性.
- 为了确定这种可穿戴设备在一般人群使用中的有效性.
主要方法:
- 招募健康参与者进行VO2max和脉冲氧计测试.
- 将Garmin fēnix 6数据与标准测量 (代谢系统,医疗级脉冲氧计) 的数据进行比较.
- 使用统计分析,包括错误和相关性分析,并使用预定义的验证标准 (CCC > 0.7,MAPE < 10%).
主要成果:
- 嘉明fēnix 6显示了准确的VO2max估计 (MAPE=7.05%,CCC=0.73对于30s平均水平).
- 该设备未能在所有条件 (CCC = 0.10) 中准确测量血液氧和 (BOS).
结论:
- 嘉明fēnix 6适用于体能环境中的VO2max估计.
- 不准确的BOS测量限制了其用于适应或疾病管理的使用.
- 独立验证对于利用可穿戴技术在健康和研究方面至关重要.
更多相关视频
10:45A Community-based Stress Management Program: Using Wearable Devices to Assess Whole Body Physiological Responses in Non-laboratory Settings
Published on: January 22, 2018
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
相关概念视频
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...