具有高分辨率功率的电子喷射漂移管离子移动性光谱仪与加热的溶解管
Marc-Aurèle Boillat1, Peter C Hauser1
1University of Basel, Department of Chemistry, Klingelbergstrasse 80, 4056, Basel, Switzerland.
Analytica chimica acta
|January 20, 2025
概括
高压漂移管离子移动性光谱学 (DTIMS) 在液体样本分析中实现了超过200个分辨率. 这种增强的DTIMS为传统分离技术提供了具有成本效益的替代方案.
科学领域:
- 分析化学 分析化学
- 物理化学 物理化学
背景情况:
- 漂流管离子移动性光谱 (DTIMS) 通常表现出有限的分离性能,分辨能力低于100.
- 提高DTIMS的解析能力需要增加漂移电压和优化离子群动态.
研究的目的:
- 开发和评估一个改进的高压DTIMS仪器,用于增强液体样本分析.
- 为了研究加热的溶解管和修改的探测器电路对分离性能的影响.
主要方法:
- 使用电喷射源构建一个高压DTIMS仪器.
- 整合加热的溶解管以优化离子转移.
- 修改探测器电路以尽量减少噪音.
- 分析具有挑战性的离子混合物,包括除草剂,聚合物和卡尼丁异构体.
主要成果:
- 对于各种离子混合物,实现了从123到228的分辨率.
- 证明了228的分辨率,代表了24.1kV的理论最大值的96%.
- 成功分离复杂混合物,包括异构体.
结论:
- 改进的DTIMS显示了分析液体样本的巨大潜力.
- 这种技术提供了一个简单的,低成本的替代色谱学和质谱学.
- 独立的电喷离子移动性光谱是未被探索的,并拥有尚未开发的分析能力.
相关概念视频
Electrospray Ionization (ESI) Mass Spectrometry
722
Higher molecular weight biomolecules are nonvolatile compounds that may decompose before ionizing or vaporizing during mass analysis with conventional electron impact ionization methods. Accordingly, electrospray ionization (ESI) is the favored method for vaporizing and ionizing biomolecules as it circumvents rapid fragmentation and enables the recording of mass signals for the entire biomolecule.
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
722
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
618
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
618
Mass Analyzers: Common Types
554
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
554
High-Resolution Mass Spectrometry (HRMS)
1.2K
The resolution of a mass spectrometer depends on the efficiency of separating ions with different ion masses. The mass of an atom is approximated to the sum of the masses of protons and neutrons inside, considering the masses of protons and neutrons as equal. However, the masses of the proton (1.6726 × 10−24 g) and neutron (1.6749 × 10−24 g) are not truly equal. There is a minor error in the expression of atomic masses relative to the simplest atom of hydrogen. For...
1.2K
Capillary Electrophoresis: Instrumentation
180
Capillary electrophoresis instrumentation typically consists of several key components. A high-voltage power supply generates the electric field necessary for the separation by connecting to an anode (the positively charged electrode) and a cathode (the negatively charged electrode) located in buffer reservoirs at each end of the capillary tube. The system includes a sample vial, a fused silica capillary tube coated with polyimide for mechanical strength through which the sample components...
180
Tandem Mass Spectrometry
886
Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and signal-to-noise ratio for the analyte. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
886