使用过氧化气体等离子体的灵活内镜的终端杀菌新技术的评估
Martin M Varghese1, Samir Memic1, Maria M Torres-Teran1
1Research Service, VA Northeast Ohio Healthcare System, Cleveland, OH, USA.
Infection control and hospital epidemiology
|January 27, 2025
概括
一种新的过氧化气体等离子体灭菌器有效地杀死了内镜上的大多数微生物,但与细菌子作斗争. 彻底的预清洗对于有效的灭菌至关重要,尤其是在高污染水平的情况下.
科学领域:
- 医疗器械灭菌医疗器械的灭菌.
- 感染控制的方法是感染控制.
- 微生物学 微生物学
背景情况:
- 内镜需要高水平的消毒或杀菌.
- 过氧化气体等离子体是一种低温灭菌方法.
- 消毒复杂的医疗器械,如内镜存在挑战.
研究的目的:
- 为了评估一种新的过氧化气体血灭菌器对内镜的有效性.
- 评估杀菌对植物生物和细菌子的有效性.
- 为了确定有机负载对灭菌性能的影响.
主要方法:
- 实验室测试了一种新的过氧化气体等离子体内镜灭菌器.
- 用高负载的植物生物和细菌子注射内镜替代体.
- 包括有机物质和盐来模拟具有挑战性的条件.
- 评估微生物减少到无法检测的水平.
主要成果:
- 实现了植物性生物的持续减少到无法检测的水平.
- 细菌子并没有一直减少到无法检测的水平.
- 绝育的有效性受到高生物负载,有机物质和盐的挑战.
结论:
- 新的过氧化气体等离子体灭菌器显示出有希望的结果,但需要进一步优化子无活化.
- 对内镜进行精心的预清洁对于有效的灭菌至关重要.
- 有机物质的存在和高的生物负担显著影响了绝育效率.
相关概念视频
Methods of Sterilization II: Chemical Methods
6.2K
In healthcare, the chemical method of sterilization uses chemical sterilants to treat surgical instruments and medical supplies to help prevent the transmission of infectious pathogens to patients. Due to heat sensitivity, most medical supplies and equipment should not be exposed to high temperatures. These parts include rubber, plastic, glass, and other similar elements.
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
6.2K
Methods of Sterilization I: Physical Methods
19.5K
As used in a healthcare facility, sterilization destroys all microorganisms through physical or chemical methods. The physical method includes steam, dry heat, boiling water, and radiation.
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
19.5K
Cleaning, Sterilization, and Disinfection
6.6K
Cleaning, disinfection, and sterilization are the methods that help to break the infection chain and prevent disease.
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
6.6K
Hand hygiene
3.1K
Asepsis is the practice of preventing or breaking the chain of infection. The nurse employs aseptic techniques to prevent the spread of microorganisms and reduce the risk of diseases. Hand hygiene is the cornerstone of aseptic techniques and is classified into medical and surgical asepsis. Medical asepsis includes hand hygiene and the use of gloves. Surgical asepsis, or the sterile technique, refers to practices that render and keep objects and areas free of microorganisms.
Hand washing...
Hand washing...
3.1K
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
502
Inductively coupled plasma (ICP) is the most widely used plasma source in atomic emission spectroscopy (AES), also known as Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES). The ICP source, or torch, consists of three concentric quartz tubes with argon gas flowing through them. A spark from a Tesla coil initiates the ionization of argon, generating a high-temperature plasma.
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
502