一个新的光谱域光学连贯性断层扫描生物仪的重复性和可重复性,以及与基于扫描源光学连贯性断层扫描生物仪的协议
Xin Li1,2,3, Chak Seng Lei1,2,3, Rui Ning1,2,3
1Eye Institute and Department of Ophthalmology and Vision Science, Eye & ENT Hospital, Fudan University, Shanghai, 200031, China.
Eye and vision (London, England)
|January 31, 2025
概括
使用光谱域光学连贯性断层扫描 (SD-OCT) 的Colombo IOL生物计显示出优异的可重复性和可重复性,用于眼睛测量. 它与IOLMaster 700非常一致,除了白至白的距离和瞳孔直径.
科学领域:
- 眼科医生 眼科 眼科
- 生物医学光学 生物医学光学
- 医疗器械 医疗器械
背景情况:
- 评价Colombo IOL生物仪,一种新的光谱域光学连贯性断层扫描 (SD-OCT) 装置.
- 评估其在正常受试者眼睛参数测量的性能.
研究的目的:
- 为了确定科伦坡IOL生物计的重复性和可重复性.
- 为了比较其与基于扫源光学连贯性断层扫描 (SS-OCT) 的IOLMaster 700.00的协议.
主要方法:
- 一项前性研究,涉及91个正常受试者的91只眼睛.
- 眼睛参数的测量,包括轴长,角膜厚度,前腔尺寸,透镜厚度,角膜测量,白,白至白的距离和瞳孔直径.
- 统计分析使用主体内部标准偏差,测试重复测试可重复性,变化系数,类内相关系数,配对的t测试和布兰德-阿尔特曼图.
主要成果:
- 科伦坡IOL生物计显示了所有测量参数的低主体内标准偏差和测试-重复测试可重复性值.
- 类内相关系数很高,表明了很好的观察者内可重复性和可重复性,一般高于0.900.
- 在大多数参数上,Colombo IOL生物计和IOLMaster 700之间观察到良好的一致性,报告了特定的一致性极限.
结论:
- 基于SD-OCT的科伦坡IOL生物计表表现出极好的重复性和可重复性.
- 该设备在大多数眼睛参数上与基于SS-OCT的IOLMaster 700有很好的一致性.
- 差异主要在白色与白色之间的距离和瞳孔直径测量方面被发现.
相关概念视频
Confocal Fluorescence Microscopy
Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
Super-resolution Fluorescence Microscopy
Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been developed.