对三种CRM进行交换性研究,评估它们适用于校准,准确性验证和测量人类血液中金属度的统计质量控制方法的适用性
Thomas P J Linsinger1, Guy Auclair2, Liesbet Deprez2
1Joint Research Centre, European Commission, Retieseweg 111, 2440, Geel, Belgium. thomas.linsinger@ec.europa.eu.
Analytical and bioanalytical chemistry
|February 4, 2025
概括
这项研究评估了血液中重金属的认证参考材料,发现目前的可交换性标准过于严格. 放松标准表明材料适用于校准器和质量控制.
科学领域:
- 分析化学 分析化学
- 环境科学 环境科学
- 临床化学 临床化学
背景情况:
- 认证参考材料 (CRM) 对于血液中准确的重金属分析至关重要.
- 可交换性确保CRM在分析过程中表现得像患者样本.
研究的目的:
- 为了评估三个血液CRM的Cd,Cr,Hg,Ni,Pb和Tl的可交换性.
- 评估非可换性偏差对测量不确定性和适合质量控制的影响.
主要方法:
- 偏差方法的差异比较消化ICP-MS与稀释ICP-MS或GFAAS.
- 最大非可交换性偏差 (MANCB) 取自能力测试提供商的容忍限度.
主要成果:
- 目前的通用性标准过于严格,需要过度测量或更大的MANCB.
- 增加的MANCB对可容忍的测量不确定性产生了小到中等的影响,表明对校准器使用的可切换性.
- 包括非可交换性不确定性在内的整体测量不确定性最小增加.
- CRM偏差通常在患者样本偏差置信限范围内,支持其在质量控制中的使用.
结论:
- 评估的CRM适用于血液重金属分析中的真实性控制,方法验证和统计质量控制.
- 调整 MANCB 标准是必要的,以便在实践中评估通用性.
- 当CRMs作为校准器使用时,它们表现出可靠的性能.
相关概念视频
Data Validation
141
Method validation is a crucial process in analytical chemistry designed to confirm that a given method consistently produces reliable and high-quality results. This process is essential when a method is applied to different sample matrices or when procedural modifications are made, ensuring that the results meet acceptable standards across various applications.
Key parameters for method validation include:
Key parameters for method validation include:
141
Instrument Calibration
149
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
149
Controlled-Current Coulometry: Overview
154
Controlled current coulometry, also known as amperostatic coulometry, is a technique used in electrochemical analysis to measure the quantity of a substance through the controlled passage of current. It involves the application of a constant current to an electrochemical cell containing the analyte of interest. As the current flows through the cell, the analyte undergoes a redox reaction at the electrode surface, resulting in a charge transfer. By monitoring the time required for a certain...
154