综合分析了FDA制造商和用户设施设备体验数据库报告的MRI相关的耳植入不良事件
Amy Lin1, Arjun K Menta2, Syed Ameen Ahmad2
1The Ohio State University College of Medicine Columbus Ohio USA.
Laryngoscope investigative otolaryngology
|February 7, 2025
概括
磁力共振成像条件下的耳植入物 (CI) 显著减少了与磁力共振成像相关的不良事件. 带有直径磁铁的设备没有报告的并发症,这表明接受MRI扫描的患者的安全性有所改善.
科学领域:
- 生物医学工程 生物医学工程
- 医疗器械安全 医疗器械安全
- 放射学 放射学是一门学科.
背景情况:
- 磁共振成像 (MRI) 对耳植入物 (CI) 接受者构成风险,原因是设备磁铁.
- 包括脱落和疼痛在内的不良事件已在MRI扫描期间与传统CI报告.
- 开发MRI条件CI的目的是减轻这些风险.
研究的目的:
- 使用FDA MAUDE数据库识别和分析与耳植入物 (CI) 相关的MRI相关不良事件.
- 为了确定具有直径磁铁的耳植入物是否与MRI相关不良事件的减少有关.
主要方法:
- 来自FDAMAUDE数据库的MRI相关不良事件报告的回顾性分析 (01/01/201301/29/2023).
- 手动提取结果,干预措施,设备问题,电极设计和患者问题.
- 使用R进行统计分析,对三大IC制造商的并发症进行分组和分析.
主要成果:
- 总共有668个与MRI相关的CI并发症被报告 (2,1%的32,446个报告).
- 常见的问题包括设备脱落/迁移 (73.3%) 和患者疼痛 (31.6%).
- 在引入MRI条件CI后,每年中位数不良事件的显著下降被观察到 (p=0.014).
结论:
- 引入MRI条件的耳植入物显然减少了MRI相关的不良事件.
- 对于来自美国三大制造商的MRI条件CI设备,没有报告MRI相关的并发症.
- 在MRI条件下的CI中,具有直径磁铁的设备似乎在MRI过程中提高了患者的安全性.
相关概念视频
The Cochlea
44.5K
The cochlea is a coiled structure in the inner ear that contains hair cells—the sensory receptors of the auditory system. Sound waves are transmitted to the cochlea by small bones attached to the eardrum called the ossicles, which vibrate the oval window that leads to the inner ear. This causes fluid in the chambers of the cochlea to move, vibrating the basilar membrane.
44.5K
Imaging Studies I: CT and MRI
175
Introduction: MRI and CT scans are crucial advancements in medical imaging techniques, playing a vital role in diagnosing conditions related to the gastrointestinal (GI) system. Each scan serves distinct purposes, targets specific areas, and requires unique nursing duties.
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
175