一个用于评估尿液流量计的协议
Alexandra Bacon1, Andrew Gammie1
1Bristol Urological Institute, Southmead Hospital, Bristol, UK.
Journal of medical engineering & technology
|February 27, 2025
概括
一个新的协议有效地验证了尿液流量计的准确性和过,用于下尿道症状调查. 这种标准化的方法确保新的尿流器件符合临床实用性和国际泌尿病咨询指南.
科学领域:
- 泌尿器科 泌尿器科 泌尿器科 泌尿器科
- 医疗设备工程 医疗设备工程
背景情况:
- 尿流量测量对于诊断下尿路症状至关重要.
- 准确可靠的尿液流量计对于有效的患者检查至关重要.
研究的目的:
- 开发和评估用于测试新型尿液流量计的标准化协议.
- 为了验证Minze Uroflow®设备的准确性和过能力.
主要方法:
- 使用恒流瓶和圆柱式流量柱进行基板测试.
- 通过复制常见的工件进行噪声过的实验室评估.
- 医疗保健专业人员在常规流动诊所中对临床可用性进行评估.
主要成果:
- 开发的协议证实了尿流参数 (Qmax和Vvoid) 在声称的准确性和ICS建议范围内.
- 敏泽Uroflow®和其他测试设备证明了有效的噪音过.
- 该协议的测试与国际泌尿病咨询指南保持一致,确保临床实用性.
结论:
- 已经建立了一个强大的协议,用于验证尿液流量计的准确性,过和可用性.
- 该协议有助于临床医生和研究人员评估新的尿液流量计模型.
- 开发的方法确保了尿液流量计器件的临床有用性.
相关概念视频
Pipe Flowrate Measurement
224
In pipe flow measurement, orifice, nozzle, and Venturi meters are commonly used to determine fluid flowrates by constricting the flow area, which increases fluid velocity and reduces pressure. This pressure difference, governed by Bernoulli's principle and adjusted for real-world conditions, is essential for calculating flowrate. Each meter type is suited to specific applications based on accuracy, efficiency, and compatibility with various flow conditions.
The orifice meter is a simple,...
The orifice meter is a simple,...
224
Measurement of Fluid Pressure
150
Fluid pressure is commonly measured using devices called manometers, which rely on liquid columns to indicate pressure differences. The height of a liquid column in a manometer reflects the pressure exerted by the fluid, providing a simple yet effective means of measurement. Different types of manometers serve specific purposes based on their configurations and the type of fluids involved.
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
150
Flow Table Test
116
The flow table test is an established method used to assess the workability of concrete, particularly useful for evaluating highly flowable concrete mixes. This test employs an apparatus that consists of a wooden board topped with a steel plate, collectively weighing 35 pounds. The board is connected to a base via a hinge and measures 27.6 inches on each side.
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
116
Assessment of blood pressure in brachial artery(one-step method)
550
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
550
Equipments Used To Measure Blood Pressure
786
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
786
Assessing Blood pressure using a doppler ultrasound
1.2K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.2K