用Tono-Vera进行眼内压力测量:一种新的手持反弹声度计
Luiz A F Beniz1,2, Alessandro A Jammal1, Davina A Malek1
1Bascom Palmer Eye Institute, University of Miami, Miami, FL.
Journal of glaucoma
|March 3, 2025
概括
新型的Tono-Vera手持度仪与眼内压力 (IOP) 测量的眼响应分析仪表达了强烈一致. 这项研究证实了Tono-Vera作为对格洛科马患者IOP评估的可靠工具.
科学领域:
- 眼科医生 眼科 眼科
- 医疗器械 医疗器械
- 玻璃眼研究研究 玻璃眼研究
背景情况:
- 眼内压力 (IOP) 是治疗青光眼的关键指标.
- 准确的IOP测量对于诊断和监测至关重要.
- 新型手持度仪旨在提高可访问性和易用性.
研究的目的:
- 为了评估Tono-Vera手持反弹音量计和眼睛反应分析仪 (ORA) 之间的协议.
- 为了评估新型带有摄像头定位系统的度仪的性能.
- 为了比较两个不同的设备的眼内压力测量.
主要方法:
- 一项涉及54只眼睛的正常眼睛和开角眼睛的玻璃眼患者的横截面研究.
- 用同一个技术人员使用Tono-Vera和ORA进行测量.
- 用线性回归模型进行校准,Bland-Altman图表评估了一致性.
主要成果:
- 校准的Tono-Vera (IOPtv) 与ORA IOPg (R2 = 0.782) 显示出强烈的相关性.
- 平均差异为-0.4 ± 2.5 mmHg,其中96.5%的测量在±5 mmHg范围内.
- 在设备之间没有观察到固定或比例偏差.
结论:
- 校准的Tono-Vera显示出与ORA IOPg有很强的相关性和一致性.
- 托诺维拉是一种可靠的手持度仪,用于测量内血压.
- 该设备显示了在临床环境中准确的IOP评估的潜力.
相关概念视频
Assessment of blood pressure in brachial artery(two-step method)
644
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
644
Open Angle Glaucoma: Treatment
383
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...
383
Special considerations while measuring blood pressure
696
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
696
Glaucoma: Overview
494
Glaucoma is an eye condition characterized by increased intraocular pressure that damages the retina and optic nerve, leading to irreversible blindness if left untreated. The human eye has various components, including the cornea, iris, pupil, lens, and optic nerve. Aqueous humor is secreted by the epithelium of the ciliary body in the posterior chamber and flows through the trabecular meshwork and canal of Schlemm, maintaining normal intraocular pressure. The trabecular meshwork and the canal...
494
Assessment of blood pressure in brachial artery(one-step method)
550
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
550
Equipments Used To Measure Blood Pressure
786
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
786