相关实验视频
Updated: May 23, 2025

10:53
A Novel Application of Musculoskeletal Ultrasound Imaging
Published on: September 17, 2013
23.3K
在手持式超声波设备中的光谱多普勒的可靠性
Dominika Filipiak-Strzecka1, Jarosław D Kasprzak2, Piotr Lipiec2
1Department and Chair of Cardiology, Bieganski Specialty Hospital, Medical University of Lodz, Kniaziewicza 1/5, Lodz, 91-347, Poland. dominika.filipiak@gmail.com.
The international journal of cardiovascular imaging
|March 10, 2025
概括
手持式超声波设备 (HUD) 提供可靠的测量,以评估透气功能和膜流速,显示高相关性和与标准心声回声学良好一致.
科学领域:
- 心脏病学 心脏病学
- 医疗成像医学成像
- 超声波技术 超声波技术 超声波技术
背景情况:
- 评估左心室透静功能和膜血液动力学对于诊断心脏病至关重要.
- 传统心声回声仪需要专门的设备和训练有素的人员,这限制了其在某些环境中的可访问性.
- 手持式超声波设备 (HUD) 为心脏评估提供了一个便携式替代方案.
研究的目的:
- 为了验证使用HUD获得的透析功能参数和门流速的准确性和可靠性.
- 为了比较HUD测量与标准的全回声心脏图检查.
主要方法:
- 一项前性研究,涉及105名因呼吸障碍或运动不耐受而转诊的患者.
- 使用HUD进行床边心声查,然后在24小时内进行参考标准全心声图.
- 分析包括相关性,布兰德-阿尔特曼图表和卡帕系数,用于扩张性参数和门速度.
主要成果:
- 在大多数测量中,在HUD和标准心声回声学之间观察到高至非常高的相关性.
- 伴心流入速度显示轻微低估偏差,而伴心环状早期透析速度 (e') 没有偏差.
- 对于主要的HFA-PEFF得分参数,发现了实质性的或几乎完美的一致性;对于加速的米特拉和大动脉流量检测,发现了良好的一致性 (分别为0.83和0.74卡帕).
结论:
- 带有多普勒功能的HUD能够可靠地评估左心室透气功能的参数.
- 精确测量大动脉和心肌峰喷流速度可以使用HUD.获得.
- HUD可以成为快速和可靠的心脏评估的宝贵工具,特别是在护理点设置中.
相关概念视频
Assessing Blood pressure using a doppler ultrasound
1.2K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.2K
Ultrasonography
4.3K
Ultrasonography is an imaging technique that uses high-frequency sound waves to visualize the body's internal structures. It is a non-invasive and safe procedure that does not involve the use of ionizing radiation, making it widely used in various medical fields. Ultrasonography is used to study heart function, blood flow in the neck or extremities, certain conditions such as gallbladder disease, and fetal growth and development.
During an ultrasonography procedure, a handheld device called...
During an ultrasonography procedure, a handheld device called...
4.3K
Equipments Used To Measure Blood Pressure
782
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
782
Doppler Effect - II
3.3K
The Doppler effect has several practical, real-world applications. For instance, meteorologists use Doppler radars to interpret weather events based on the Doppler effect. Typically, a transmitter emits radio waves at a specific frequency toward the sky from a weather station. The radio waves bounce off the clouds and precipitation and travel back to the weather station. The radio frequency of the waves reflected back to the station appears to decrease if the clouds or precipitation are moving...
3.3K
Ultrasound II: Endoscopic Ultrasound and FibroScan
63
Endoscopic Ultrasound (EUS) and FibroScan are valuable diagnostic tools in gastroenterology and hepatology, each with specific applications and techniques.
Endoscopic Ultrasound (EUS):
Endoscopic Ultrasound (EUS):
63
UV–Vis Spectrometers
1.2K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
1.2K