极度早产新生儿的非侵入性血压测量:不同方法的比较分析
Shalabh Garg1,2, Adnan Zafar3
1Department of Neonatology, The James Cook University Hospital, Middlesbrough, UK. s.garg@nhs.net.
Indian pediatrics
|April 1, 2025
概括
使用多普勒和振荡计方法对极度早产新生儿的非侵入性血压监测显示不准确. 与侵袭性动脉导管测量相比,振荡计读数比多普勒更可靠.
科学领域:
- 新生儿医学 新生儿医学
- 儿科心脏病学 儿科心脏病学
- 医疗器械技术 医疗器械技术
背景情况:
- 精确的血压监测对于极其早产新生儿至关重要.
- 侵入性方法,如动脉导管 (UAC),提供可靠的数据,但存在风险.
- 非侵入性方法是可取的,但它们在这个人群中的准确性是有争议的.
研究的目的:
- 为了比较非侵入性多普勒和振荡计血压测量与侵入性UAC读数在怀孕28周以下的新生儿的准确性.
- 为了确定哪种非侵入性方法在这个脆弱群体中为血压监测提供了更高的可靠性.
主要方法:
- 一项涉及新生儿<28周妊娠期需要UAC插入的前性研究.
- 使用侵入性UAC,非侵入性振荡计和多普勒方法同时测量血压.
- 分析了11名新生儿的438个血压记录,每名婴儿有10-15个侵入性读数.
主要成果:
- 与UAC相比,这两种非侵入性方法都显示出变异性,低估和高估血压.
- 振荡计法比多普勒法 (17%) 更频繁地 (41%) 达到UAC值10%内的读数.
- 显著的高估和低估发生在两个非侵入性技术.
结论:
- 极度早产新生儿的非侵入性血压监测容易出现不准确性,与侵入性UAC相比,低估和高估值.
- 在这种人群中,振荡计法在非侵入性血压评估中表现出高于多普勒法的可靠性.
- 可能需要进一步的研究来完善最佳新生儿护理的非侵入性技术.
相关概念视频
Equipments Used To Measure Blood Pressure
762
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
762
Sites for measruring blood pressure
1.4K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
1.4K
Assessment of blood pressure in brachial artery(two-step method)
616
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
616
Special considerations while measuring blood pressure
689
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
689
Measurement of Blood Pressure
610
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
610
Assessment of blood pressure in brachial artery(one-step method)
544
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
544