Adrian G Dumitrascu1, Michelle Perry2, Ana-Maria Chindris2
1Division of Hospital Internal Medicine, Mayo Clinic Florida, Jacksonville, Florida, USA.
Diabetes technology & therapeutics
|April 14, 2025
概括
患者的个人持续血糖监测器 (CGM) 显示出足够的医院准确性,除了低血糖症. 生命体征和实验室值没有影响CGM的表现,这表明可靠的住院血糖监测.
科学领域:
- 内分泌学 在内分泌学.
- 医疗器械 医疗器械
- 糖尿病管理 糖尿病管理
背景情况:
- 连续血糖监测器 (CGM) 越来越多地用于糖尿病管理.
- 在住院治疗中评估CGM准确性对于患者安全和有效的血糖控制至关重要.
研究的目的:
- 评估患者在医院环境中使用的个人CGM的准确性.
- 确定CGM准确度与实验室葡萄糖值以及生命体征之间的相关性.
- 将住院患者的CGM性能指标与已发表的文献和临床指南进行比较.
主要方法:
- 在入院时使用个人CGM的糖尿病成年患者被纳入非临界护理环境.
- CGM读数与照顾点 (POC) 葡萄糖计和实验室 (实验室) 血糖测量时间配对.
- 使用平均绝对相对差异 (MARD) 和克拉克错误网格来评估准确性;对生命体征和实验室值进行了相关性分析.
主要成果:
- CGM-POC对产生了13.7%的MARD (Dexcom: 14.4%,Libre: 11.8%),而CGM-Lab对显示了13.6%的MARD (Dexcom: 12.7%,Libre: 15.4%). 这两种组的MARD值均为1.3%.
- 在克拉克错误网格上,超过98%的CGM-POC和CGM-Lab对位于可接受区域 (A和B) 之内.
- 没有发现CGM准确度与平均动脉血压,血红蛋白,淋巴膜过率或脉冲氧计之间的显著相关性.
结论:
- 个人CGM显示出足够的准确性用于住院血糖监测,在低血糖范围中指出了限制.
- 住院患者的CGM性能不受生命体征或实验室结果的显著影响.
- 观察到的时间范围 (TIR) 和葡萄糖可变度指标与住院患者的指导方针保持一致,并且与医院插入的传感器可比.
相关概念视频
Pulse rhythm
721
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
721
Hypoglycemia and Glucagon
142
Without prolonged fasting, healthy individuals maintain blood glucose levels above 3.5 mM due to a well-adapted neuroendocrine counterregulatory system that effectively prevents acute hypoglycemia, a potentially life-threatening condition. The primary clinical scenarios for hypoglycemia encompass diabetes treatment, inappropriate production of endogenous insulin or insulin-like substances by tumors, and the use of glucose-lowering agents in non-diabetic individuals. Notably, hypoglycemia in the...
142
Errors occurring during blood pressure monitoring
568
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
568
SBAR II: Application of SBAR
4.2K
SBAR is an effective communication tool used by healthcare professionals to communicate patient information accurately. SBAR stands for Situation, Background, Assessment, and Recommendation. For a better understanding, an example is given below.
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
4.2K