磁化转移效应对定量MR成像中的T1测量的影响:比较二维多动态多回声 (2D-MDME) 和三维量化,使用交叉的Look-Locker获取序列 (3D-QALAS) 协议对两个系统
Ryoichi Kose1, Katsumi Kose1, Yasuhiko Terada2
1MRIsimulations, Tokyo, Japan.
概括
磁化转移效应显著影响2D多动态多重回声 (2D-MDME) MRI中的T1测量,导致缩短. 然而,3D-QALAS (3D-QuAntification使用交叉的Look-Locker获取序列与T2准备脉冲) 没有受到影响,这凸显了在多参数定量MRI比较中需要考虑MT效应的必要性.
科学领域:
- 磁共振成像技术 磁共振成像技术
- 量化MRI是指数量化的MRI.
- 生物医学工程 生物医学工程
背景情况:
- 多参数定量核磁共振 (MP-qMRI) 方法可同时测量多个放松时间.
- 不同的MP-qMRI技术之间的关系仍然不清楚,阻碍了直接比较.
- 2D多动态多重回声 (2D-MDME) 是广泛使用的,但与3D-QALAS (3D-QuAntification使用交联的Look-Locker获取序列与T2准备脉冲) 等潜在的继任者进行比较是缺乏的.
研究的目的:
- 为了研究2D-MDME和3D-QALAS MP-qMRI方法之间的关系.
- 评估磁化转移 (MT) 影响对两种技术T1测量的影响.
- 确定2D-MDME和3D-QALAS在临床环境中的可比性.
主要方法:
- 在1.5T和3TMRI系统上实施2D-MDME和3D-QALAS.
- 使用MP-qMRI两种方法测量幻影的放松时间.
- 对MT影响对T1放松时间测量的影响进行了集中分析.
主要成果:
- 由于MT的影响,2D-MDME显示T1缩短约20%-35%.
- 3D-QALAS显示MT对T1测量的影响没有显著的影响.
- 这两种方法之间的T1测量差异归因于MT效应.
结论:
- 在临床研究中比较2D-MDME和3D-QALAS时,必须考虑磁化转移效应.
- 对于MT效应,3D-QALAS似乎比2D-MDME对T1量化更强大.
- 仔细的方法选择和数据解释对于准确的MP-qMRI分析至关重要.
相关概念视频
Magnetic Resonance Imaging
4.8K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
4.8K
Two-Dimensional (2D) NMR: Overview
553
The 1D NMR spectrum of large and complex molecules like natural products has complicated splitting patterns and overlapping signals, which can be easily interpreted using 2-dimensional (2D) NMR. Unlike 1D NMR, 2D NMR has two frequency axes that provide the coupling information between the nucleus A and nucleus B in a molecule. The process from which 2D spectra are obtained has four steps.
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....
553
Atomic Nuclei: Magnetic Resonance
584
The number of nuclear spins aligned in the lower energy state is slightly greater than those in the higher energy state. In the presence of an external magnetic field, as the spins precess at the Larmor frequency, the excess population results in a net magnetization oriented along the z axis. When a pulse or a short burst of radio waves at the Larmor frequency is applied along the x axis, the coupling of frequencies causes resonance and flips the nuclear spins of the excess population from the...
584