根据动脉协会的建议,对基于袖口的中央血压装置的验证评估
Tan V Bui1, Dean S Picone1,2, Wojciech P Kosmala1,3
1Menzies Institute for Medical Research, University of Tasmania, Hobart.
Blood pressure monitoring
|May 6, 2025
概括
针对非侵入性中央血压估计的Uscom BP+设备,与侵入性测量相比,不符合准确性标准. 这种基于袖子的设备未能通过临床使用的标准.
科学领域:
- 心血管研究的心血管研究.
- 生物医学工程 生物医学工程
- 医疗器械验证 医疗器械验证
背景情况:
- 中心血压 (BP) 估计提供了超出手臂血压的洞察力.
- 乌斯康 BP+ 是一款基于袖口的设备,旨在进行非侵入性的中央血压测量.
- 准确的中央血压监测对于评估心血管风险至关重要.
研究的目的:
- 为了评估Uscom BP+设备用于估计中心BP的准确性.
- 为了比较来自Uscom BP+的非侵入性中心血压读数与侵入性测量.
- 为了确定Uscom BP+设备是否符合动脉协会准确性标准.
主要方法:
- 在191名参与者中同时记录非侵入性袖口中央血压 (Uscom BP+) 和侵入性中央大动脉血压.
- 参与者在三个国际研究站点进行了冠状动脉血管造影.
- 验证遵循了动脉学会的建议,通过标准为≤5 ± ≤8 mmHg.
主要成果:
- 使用默认校准,Uscom BP+显著低估了缩血压 (-10.2 ± 11.2 mmHg) 和高估了透缩血压 (9.8 ± 8.5 mmHg).
- 替代校准方法稍微提高了缩血压的准确性,但保持了很高的可变性.
- 透析血压的准确性和可变性在所有校准方法中保持相似.
结论:
- 美国电信公司的BP+袖口装置未能满足动脉协会对中央血压测量的准确性标准.
- 目前的校准方法无法确保Uscom BP+设备在临床应用中具有足够的准确性.
- 可能需要进一步的研究来提高非侵入性中央血压估计设备的准确性.
相关概念视频
Pre-Procedural Guidelines for Assessing Blood Pressure
509
Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
509
Sites for measruring blood pressure
1.4K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
1.4K
Assessment of blood pressure in brachial artery(two-step method)
616
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
616
Assessment of blood pressure in brachial artery(one-step method)
544
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
544
Special considerations while measuring blood pressure
689
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
689
Equipments Used To Measure Blood Pressure
762
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
762