在酒精饮料中通过稀释和射击分子排放光谱法与溶液阳极发光放电激发源的快速乙醇测定
Krzysztof Gręda1, Klaudia Kowalczyk1, Monika Zabłocka-Malicka1
1Wroclaw University of Science and Technology, Faculty of Chemistry, Division of Analytical Chemistry and Chemical Metallurgy, Wyspianskiego 27, 50-370 Wrocław, Poland.
Food chemistry
|June 7, 2025
概括
使用分子排放光谱法 (MES) 的新方法可以快速准确地分析饮料中的乙醇. 这种高通量技术简化了样品制备,并准确量化了乙醇含量.
科学领域:
- 分析化学 分析化学
- 频谱学是一种光谱学.
背景情况:
- 精确的乙醇测定对于饮料行业的质量控制至关重要.
- 传统方法通常需要复杂的样本准备,限制吞吐量.
研究的目的:
- 开发一种高通量分子排放光谱法 (MES) 方法,用于快速确定乙醇.
- 为了评估方法的精度,检测极限,线性和矩阵效应阻力.
主要方法:
- 样品被稀释并直接引入溶液阳极发光放电 (SAGD) 系统.
- 通过在206.8nm处测量CO分子带来量化乙醇.
- 一种稀释和射击方法实现了每小时50个样本的分析吞吐量.
主要成果:
- 获得了高精度 (1-3%) 和低检测极限 (0.005%v/v).
- 观察到高达0.6%v/v的优异线性和对矩阵效应的强烈抗性.
- 对各种酒精饮料 (相对误差 -1.7%至+7.0%) 确认了与氧化还原定位的良好一致性.
结论:
- 开发的SAGD MES方法为酒精饮料中乙醇的确定提供了快速,精确和准确的方法.
- 该方法消除了样品蒸的需要,并显示了分析其他挥发性有机化合物的潜力.
相关概念视频
Flame Photometry: Lab
377
In a flame photometer, when a solution like potassium chloride is aspirated into the flame, the solvent evaporates, leaving behind dehydrated salt. This salt dissociates into free gaseous atoms in their ground state. Some of these atoms absorb energy from the flame, leading to their excitation. The excited atoms return to the ground state, emitting photons at characteristic wavelengths. Because only electronic transitions are involved, the resulting emission lines are very narrow. The intensity...
377
Atomic Emission Spectroscopy: Instrumentation
629
The instrumentation of atomic emission spectrometry (AES) involves various components, including atomization devices that convert samples into gas-phase atoms and ions. There are two main types of atomization devices: continuous and discrete atomizers. Continuous atomizers, like plasmas and flames, introduce samples in a constant stream, while discrete atomizers inject individual samples using syringes or autosamplers. The most common discrete atomizer is the electrothermal atomizer.
629
Atomic Emission Spectroscopy: Overview
2.6K
Atomic emission spectroscopy (AES) is an analytical technique used to determine the elemental composition of a sample by analyzing the light emitted from excited atoms. In AES, atoms in a sample are excited to higher energy levels by thermal energy from high-temperature sources, such as plasma, arcs, or sparks. When these excited atoms return to lower energy states, they emit light at specific wavelengths characteristic of each element. The resulting atomic emission spectrum, which consists of...
2.6K
Atomic Emission Spectroscopy: Lab
252
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
252
Atomic Absorption Spectroscopy: Atomization Methods
674
Atomic Absorption Spectroscopy (AAS) atomizes samples through flame atomization or electrothermal atomization. Flame atomization typically involves a nebulizer and spray chamber assembly to combine the sample with a fuel–oxidant mixture, creating a fine aerosol mist that enters a burner. Typically, the fuel and oxidant are combined in an approximately stoichiometric ratio. However, for atoms that are easily oxidized, a fuel-rich mixture may be more advantageous. Only about 5% of the...
674
Atomic Emission Spectroscopy: Interference
285
In atomic emission spectroscopy (AES), high-temperature atomizers excite a broad range of elements and molecules that generate complex emissions from sources such as oxides, hydroxides, and flame combustion products in the flame or plasma. Several strategies can be employed to minimize spectral interferences caused by overlapping emission lines or bands. These include increasing instrument resolution, choosing alternative emission lines, optimally placing the detector in low-background regions,...
285