LIBSLIBS.

Jiayuan Zhou1, Weihua Huang1, Harse Sattar2

  • 1Wuhan National Laboratory for Optoelectronics (WNLO), Huazhong University of Science and Technology, Wuhan, Hubei, 430074, China.

Talanta
|June 24, 2025
PubMed
概括

这项研究引入了声光光谱融合激光诱导分解光谱学 (AOSF-LIBS),以克服元素分析中的矩阵效应. 这种新方法显著提高了用于精确定量测量的光谱精度.

相关概念视频

Doppler Effect - II01:05

Doppler Effect - II

The Doppler effect has several practical, real-world applications. For instance, meteorologists use Doppler radars to interpret weather events based on the Doppler effect. Typically, a transmitter emits radio waves at a specific frequency toward the sky from a weather station. The radio waves bounce off the clouds and precipitation and travel back to the weather station. The radio frequency of the waves reflected back to the station appears to decrease if the clouds or precipitation are moving...
3.6K
Doppler Effect - I00:56

Doppler Effect - I

The Doppler effect and Doppler shift were named after the Austrian physicist and mathematician Christian Johann Doppler in 1842, who conducted experiments with both moving sources and moving observers. Consider an observer standing on a street corner, observing an ambulance with a siren sound passing by at a constant speed. The observer experiences two characteristic changes in the sound of the siren. Initially, the sound increases in loudness as the ambulance approaches and decreases in...
3.8K
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations01:08

IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations

Identical bonds within a polyatomic group can stretch symmetrically (in-phase) or asymmetrically (out-of-phase). Similar to hydrogen bonding, these vibrations also influence the shape of the IR peak. Generally, asymmetric stretching frequencies are higher than symmetric stretching frequencies. For example, primary amines exhibit two distinct IR peaks between 3300–3500 cm−1 corresponding to the symmetric and asymmetric N-H stretching, while secondary amines exhibit a single...
1.2K
Distance Corrections01:15

Distance Corrections

To achieve precise distance measurements, especially in surveying and construction, certain corrections must be applied to account for potential sources of error like the standardization errors, temperature variations, and slope adjustments.Standardization error emerges when measurement equipment undergoes changes, such as wear, repairs, or weather impacts. To address this, surveyors compare the equipment’s readings to a standard. This process identifies any deviation that might lead to...
92
NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
784
Instrument Calibration01:12

Instrument Calibration

Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
281