第四代FloTrac软件脉冲轮分析用于测量和趋势心脏输出:一种方法比较研究
Caroline Dinesen1,2, Simon Tilma Vistisen2, Rasmus Aagaard1,2,3
1Department of Cardiothoracic and Vascular Surgery, Anesthesia Section, Aarhus University Hospital, Aarhus, Denmark.
由于不可接受的百分比错误和有限的趋势能力,FloTrac方法,一种微创心脏输出监测器,在心脏手术后的患者中不能与热稀释方法互换.
科学领域:
- 心血管生理学心血管生理学
- 医疗器械评估 医疗器械评估
- 血液动力学监测 血液动力学监测
背景情况:
- 准确的心脏输出 (CO) 评估对于血液动力学管理至关重要.
- 在FloTrac系统提供最少侵入性,连续的CO估计.
- 这项研究将FloTrac方法与肺动脉导管热稀释进行了比较.
研究的目的:
- 为了评估FloTrac方法的准确性,精度和趋势性能.
- 为了比较FloTrac的CO测量与热稀释 (TD) 的CO测量.
- 确定FloTrac与TD在心脏手术患者中的可互换性.
主要方法:
- 对41名心脏手术患者进行前性干预研究.
- 使用FloTrac和TD在预加载干预前/后同时进行CO测量.
- 分析偏差,一致限度,百分比错误 (PE) 和一致率.
主要成果:
- 偏差在 -0.36 到 0.79 L/min 之间;一致性极限很宽.
- 百分比错误 (PE) 超过了30%的门 (35-42%).
- 符合率低于90%的标准 (72-85%).
结论:
- 该FloTrac方法没有满足与热稀释的可互换性标准.
- 无法接受的百分比错误和不良的趋势限制了FloTrac的临床效用.
- 对于这种患者群体,FloTrac的准确性和精度是不够的.
更多相关视频
11:04Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism
Published on: September 1, 2014
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
相关概念视频
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...