用二极管探测器阵列测量辐射场质量中心来确定光束延迟的新方法
Mathias Dierl1, Ndimofor Chofor2, Arnd Röser3
1Klinikum Bayreuth GmbH, Department for Medical Physics, Bayreuth, Germany; Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany; Comprehensive Cancer Center Erlangen-EMN (CCC ER-EMN), Erlangen, Germany.
Zeitschrift fur medizinische Physik
|June 25, 2025
概括
这项研究开发了一种使用二极管探测器阵列和表面导向放射治疗 (SGRT) 系统测量放射治疗束延迟的新方法. 结果显示,在不同的线性加速器供应商和束参数之间,光束开启/关闭延迟的显著变化.
科学领域:
- 医学物理 医学物理
- 放射治疗技术 放射治疗技术
- 辐射瘤学 辐射瘤学
背景情况:
- 精确测量beam-on/off延迟对于精确放射治疗至关重要.
- 表面导向放射疗法 (SGRT) 系统通过跟踪患者运动来提高治疗准确性.
- 现有的延迟测量方法通常需要多个设备的同步,增加复杂性.
研究的目的:
- 提出一种新的,同步的方法来测量SGRT系统中的光束开/关延迟.
- 用单个时间测量与二极管探测器阵列来评估延迟.
- 为了评估线性加速器 (linac) 供应商,剂量速率和光束类型对延迟的影响.
主要方法:
- 使用了与Elekta Versa HD和Varian TrueBeam联机集成的AlignRT SGRT系统.
- 在运动平台上使用二极管探测器阵列,同时捕获运动和辐射传递.
- 通过将剂量分布的质量中心 (CoM) 与辐射输送时间线映射而计算的延迟.
- 在各种剂量速率和运动模式下,评估了6 MV平面化 (FF) 和平面化无波器 (FFF) 束.
主要成果:
- 梁线延迟平均为Versa HD (6 MV FFF) 的1650 ± 110毫秒,TrueBeam的280 ± 90毫秒.
- 维萨HD光束的延迟与FF光束的剂量率有所不同,较低的速度增加.
- 射线关闭延迟平均为230 ± 90毫秒 (Versa HD) 和360 ± 90毫秒 (TrueBeam),基本上与剂量速率无关.
- 在 linac 供应商和光束配置之间观察到延迟差异.
结论:
- 该 CoM 方法有效量化使用集成的 SGRT 和二极管阵列测量波束开/关延迟.
- 延迟受到 linac 供应商,剂量速率和光束能量的影响.
- 该方法提供了对延迟的高效评估,有助于定义在立体射线手术 (SRS) 中提高精度和安全性的目标边缘.
相关概念视频
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Electronic Distance Measuring Instruments
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over short distances...