使.

Ghaith Al-Refai1, Hisham Elmoaqet1, Abdullah Al-Refai2

  • 1Department of Mechatronics Engineering, German Jordanian University, Amman, Jordan.

PubMed
概括

这项研究引入了在低光图像中对象检测的两阶段系统. 该系统通过使用深度学习首先提高图像质量,然后再应用对象检测算法来显著提高检测准确性.

相关概念视频

Light Acquisition02:16

Light Acquisition

In order to produce glucose, plants need to capture sufficient light energy. Many modern plants have evolved leaves specialized for light acquisition. Leaves can be only millimeters in width or tens of meters wide, depending on the environment. Due to competition for sunlight, evolution has driven the evolution of increasingly larger leaves and taller plants, to avoid shading by their neighbors with contaminant elaboration of root architecture and mechanisms to transport water and nutrients.
8.6K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
7.1K
Deconvolution01:20

Deconvolution

Deconvolution, also known as inverse filtering, is the process of extracting the impulse response from known input and output signals. This technique is vital in scenarios where the system's characteristics are unknown, and they must be inferred from the observable signals.
Deconvolution involves several mathematical techniques to derive the impulse response. One common approach is polynomial division. In this method, the input and output sequences are treated as coefficients of...
263
Phase Contrast and Differential Interference Contrast Microscopy01:26

Phase Contrast and Differential Interference Contrast Microscopy

Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
9.8K
Super-resolution Fluorescence Microscopy01:37

Super-resolution Fluorescence Microscopy

Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
7.9K