在密集血液护理期间持续监测葡萄糖的真实准确性:一项前性研究
Marieke Tienstra1, Riemer A Been2, Janneke W de Boer1
1Department of Hematology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
Diabetes technology & therapeutics
|July 2, 2025
概括
持续血糖监测 (CGM) 对于接受密集血液学治疗的患者来说是可行的,尽管存在一些准确性限制. 这项技术可以帮助管理高血糖,而不会造成有害的胰岛素剂量错误.
科学领域:
- 内分泌学 在内分泌学.
- 血液学 血液学 血液学
- 医疗技术 医疗技术 医学技术
背景情况:
- 血液性恶性瘤患者容易患上高血糖症,影响治疗结果.
- 持续的葡萄糖监测 (CGM) 显示出有希望的毛细血管方法检测高血糖.
- 在密集的血液学治疗过程中,CGM的准确性仍在调查中.
研究的目的:
- 为了评估CGM的准确性与毛细血管点治疗 (POC) 葡萄糖测量相比.
- 评估在接受密集血液学护理的成年患者中CGM的性能.
- 在仿真抗原受体T细胞治疗,全原干细胞移植 (allo-SCT) 和自身干细胞移植 (auto-SCT) 期间调查CGM的可行性.
主要方法:
- 一项前性研究,比较了同时进行的CGM和POC葡萄糖测量.
- 包括60名接受密集血液学治疗的成年患者.
- 使用平均绝对相对差 (MARD),糖尿病技术协会 (DTS) 错误网格分析和百分比协议的准确性评估.
主要成果:
- 总体平均绝对相对差异 (MARD) 为21.5%,在治疗类型 (allo-SCT 18.3%,自动SCT 27.1%) 中存在差异.
- 葡萄糖值在15毫克/分升范围内达到了38.1%,在20毫克/分升范围内达到了51.1%,在30毫克/分升范围内达到了75.1%.
- 糖尿病技术协会 (DTS) 错误网分析表明临床准确性很好,在A+B区有99.6%的对.
结论:
- 尽管MARD相对较高,但CGM在密集血液学护理中表现出良好的临床准确性.
- 在这种患者群体中,使用CGM不太可能导致显著的胰岛素剂量错误.
- 持续的葡萄糖监测似乎是可行的和安全的管理高血糖在密集的血液学治疗期间.
相关概念视频
Pulse rhythm
937
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
937
Errors occurring during blood pressure monitoring
983
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
983
Hypoglycemia and Glucagon
351
Without prolonged fasting, healthy individuals maintain blood glucose levels above 3.5 mM due to a well-adapted neuroendocrine counterregulatory system that effectively prevents acute hypoglycemia, a potentially life-threatening condition. The primary clinical scenarios for hypoglycemia encompass diabetes treatment, inappropriate production of endogenous insulin or insulin-like substances by tumors, and the use of glucose-lowering agents in non-diabetic individuals. Notably, hypoglycemia in the...
351
Continuous Renal Replacement Therapy
156
Continuous Renal Replacement Therapy, also known as CRRT, is a procedural treatment for acute kidney injury (AKI) that gradually removes uremic toxins and fluids while maintaining acid-base balance and stabilizing electrolytes. It is particularly useful for hemodynamically unstable patients. Unlike intermittent hemodialysis, which is faster, CRRT provides a gentler approach over 24 hours, closely mimicking the function of natural kidneys. However, CRRT is not ideal for patients with...
156