在心率快速变化期间,蓝牙可穿戴心率监测器的动态响应
Mariah Sabioni1, Jonas Willén1, Seraina A Dual1
1Department of Biomedical Engineering and Health Systems, KTH Royal Institute of Technology, Stockholm, Sweden.
Physiological measurement
|July 7, 2025
概括
商用心率监测器显示对快速心率变化的响应时间有所不同. Movesense active 显示出最快的 RRI 延迟,而所有设备在强度炼期间都难以准确捕捉峰值和低谷.
科学领域:
- 生物医学工程 生物医学工程
- 可穿戴技术可穿戴技术
- 心血管生理学心血管生理学
背景情况:
- 可穿戴心率 (HR) 监测器被广泛用于健身和健康跟踪.
- 它们在动态HR变化期间的准确性,对于运动生理学至关重要,往往没有得到验证.
- 现有的验证研究经常忽视这些设备的动态响应特性.
研究的目的:
- 量化和评估商用蓝牙胸部戴式HR监视器RR间隔 (RRI) 和HR测量的动态响应.
- 在诱导快速HR变化时评估设备性能,模拟各种炼场景.
- 在不同的可穿戴显示器上比较RRI和HR数据的延迟和一致性.
主要方法:
- 合成心电图 (ECG) 信号被生成以模拟快速的HR变化 (步骤,运动,间歇性运动).
- 四个商用蓝牙胸部戴式HR监控器 (Garmin HRM-Dual,Movesense active,Polar H10,Wahoo TRACKR) 进行了测试.
- 记录了RRI和HR数据,并根据参考信号评估了延迟 (RRI和HR) 和一致性.
主要成果:
- RRI延迟差异很大,Movesense active (0.4s) 显示出最快的响应,而Polar H10 (2.6s) 和Wahoo TRACKR (2.1s) 则较慢.
- 在设备和测试条件之间,HR响应延迟也存在差异;Movesense active始终表现出最低的HR延迟.
- 所有测试的设备在间歇性炼期间都低估了HR峰值和高估了HR谷,在较低的HR值中观察到更大的不准确性.
结论:
- 商业人力资源监控器采用各种数字过技术,影响其准确捕捉快速人力资源波动的能力.
- RRI和HR测量的动态特征至关重要,在标准验证协议中经常被忽视.
- 对于需要精确跟踪急剧HR变化的应用 (例如,高强度间歇训练),建议采用节拍对节拍RRI记录,而不是处理HR数据.
更多相关视频
相关概念视频
Factors Influencing Heart Rate
4.8K
The heart rate, or pulse rate, is a vital indicator of cardiovascular health. It reflects the number of times the heart beats per minute. Various physiological and environmental factors influence heart rate, increasing or decreasing cardiac output. Understanding these factors is crucial for assessing heart function and identifying potential health issues.
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
4.8K
Pulse rhythm
934
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
934
Regulation of Heart Rates
2.2K
The regulation of heart rate is a complex process controlled by the autonomic nervous system (ANS), hormonal influences, and intrinsic cardiac mechanisms. The ANS has two main components: the sympathetic nervous system (SNS) and the parasympathetic nervous system (PNS).
The SNS increases heart rate through the release of norepinephrine and epinephrine, which act on beta-1 adrenergic receptors in the heart. This action increases the rate of depolarization in the sinoatrial (SA) node, the heart's...
The SNS increases heart rate through the release of norepinephrine and epinephrine, which act on beta-1 adrenergic receptors in the heart. This action increases the rate of depolarization in the sinoatrial (SA) node, the heart's...
2.2K
Exercise and Cardiovascular Response
1.1K
Exercise significantly impacts cardiovascular response, which is crucial for understanding patient health and designing effective treatment plans.
Light to moderate physical activity initiates a series of interconnected responses in the body. The heart rate modestly increases in anticipation of the workout, followed by widespread vasodilation as oxygen consumption by skeletal muscles increases. This results in decreased peripheral resistance, increased capillary blood flow, and accelerated...
Light to moderate physical activity initiates a series of interconnected responses in the body. The heart rate modestly increases in anticipation of the workout, followed by widespread vasodilation as oxygen consumption by skeletal muscles increases. This results in decreased peripheral resistance, increased capillary blood flow, and accelerated...
1.1K
Cardiac Output I:Effect of Heart Rate on Cardiac Output
1.3K
Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
1.3K
Equipments Used To Measure Blood Pressure
2.5K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.5K