10米的Rowland光谱仪在MAX IV Veritas光线线上进行了对接
Marcus Agåker1, Carl Johan Englund2, Victor Ekholm1
1MAX IV Laboratory, Lund University, PO Box 118, SE-22100 Lund, Sweden.
Journal of synchrotron radiation
|July 23, 2025
概括
一个新的软X射线光谱仪增强了共振无弹性散射实验. 这种仪器为先进材料研究提供了高分辨率和时间解析能力.
科学领域:
- 材料科学 材料科学 材料科学
- 原子和分子物理 原子和分子物理
- 频谱学是一种光谱学.
背景情况:
- 响应无弹性软X射线散射 (RIXS) 是研究材料电子和磁性质的强大技术.
- 高分辨率和时间分辨率测量对于理解材料中的动态过程至关重要.
研究的目的:
- 为RIXS实验推出一款新的牧场发生率罗兰谱仪.
- 为了在Veritas光束线上实现高光谱和时间分辨率的测量.
主要方法:
- 设计和实施一个带有配合元件的牧场发生率罗兰谱仪.
- 使用延迟线探测器进行时间解析测量和时间间隔.
- 在动量转移研究中采用旋转能力.
主要成果:
- 该光谱仪在250-950 eV范围内运行,分辨率>30,000在第一阶段.
- 时间分辨率为270ps,可以进行时间分辨率的实验,并改善信号噪声比.
- 通过光谱,射线追踪,振动和运动测量成功进行评估.
结论:
- 开发的光谱仪是高级RIXS研究的宝贵工具.
- 它的功能使能量,动量和分散光子的极化进行了详细的描述.
- 该仪器已准备好安装极相仪,可进行全面的光子分析.
相关概念视频
UV–Vis Spectrometers
1.5K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
1.5K
IR Spectrometers
1.5K
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
1.5K
Raman Spectroscopy Instrumentation: Overview
534
A conventional Raman spectrophotometer includes a laser source, a sample holding system, a wavelength selector, and a detector.
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
534
Atomic Absorption Spectroscopy: Instrumentation
975
An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.
The atomizer used in AAS can be either a flame atomizer or an...
The atomizer used in AAS can be either a flame atomizer or an...
975
Atomic Emission Spectroscopy: Instrumentation
607
The instrumentation of atomic emission spectrometry (AES) involves various components, including atomization devices that convert samples into gas-phase atoms and ions. There are two main types of atomization devices: continuous and discrete atomizers. Continuous atomizers, like plasmas and flames, introduce samples in a constant stream, while discrete atomizers inject individual samples using syringes or autosamplers. The most common discrete atomizer is the electrothermal atomizer.
607
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
299
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
299