在PET/CT检查中,Safire CT代重建协议可降低患者有效剂量
Alice Ferretti1, Sotirios Chondrogiannis2, Jlenia Peretto2
1Medical Physics Unit, Santa Maria della Misericordia Hospital of Rovigo, ULSS5 Polesana, Rovigo, Italy.
概括
在3级的Siemens SAFIRE代重建可以在18F-FDG PET/CT扫描中将CT剂量降低高达40%,而不会影响图像质量或SUV精度. 这种方法显著降低了患者的辐射暴露,从10.1mSv降至8.3mSv.
科学领域:
- 医疗成像医学成像
- 辐射物理学 辐射物理学
- 核医学是一种核医学.
背景情况:
- 标准过后投影 (FBP) 在PET/CT中CT重建限制剂量降低.
- 代重建技术可以降低剂量,同时保持图像质量.
研究的目的:
- 实施和评估CT代重建 (SAFIRE) 以减少18F-FDG PET/CT的剂量.
- 评估SAFIRE对图像质量和标准化吸收值 (SUV) 准确性的影响.
主要方法:
- 使用Catphan 600和Alderson-Rando幻影来测试SAFIRE在降低mA设置的情况下.
- 使用NEMA 2012年智商幻影评估PET的定量准确性.
- 对25名患者进行临床评估,将SAFIRE与FBP进行比较.
主要成果:
- SAFIRE 3级降低CT剂量~40% (幻影) 和30% (人形幻影) 没有降低噪音或CNR.
- 在剂量优化的代重建中,PET SUV值显示出最小的偏差 (<0.6%).
- 临床评估证实,与FBP相比,SAFIRE降低剂量的图像质量不劣.
结论:
- 在3级的SAFIRE代重建有效地将全身PET/CT中平均有效剂量降低了1.8mSv.
- 这种剂量降低可以实现,而不会损害图像质量或SUV精度.
- 在PET/CT检查中,SAFIRE提供了一种可行的策略来降低患者的辐射暴露.
相关概念视频
Imaging Studies III: Computed Tomography
52
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
52
Imaging Studies for Cardiovascular System V: CT
66
Cardiac computed tomography (CT) scanning is an advanced cardiac imaging technique that utilizes CT technology, with or without intravenous (IV) contrast, to produce accurate cross-sectional virtual slices of specific areas of the heart, coronary circulation, and major blood vessels such as the aorta, pulmonary veins, and arteries. The computer processes these slices to generate three-dimensional images. Multidetector CT (MDCT) is a rapid form of CT scanning that captures multiple slices...
66
Radiological Investigation III: Pulmonary Angiogram and PET Scan
170
Radiological investigations are paramount in the diagnosis and management of various pulmonary diseases. Two essential investigations are the Pulmonary Angiogram and the Positron Emission Tomography (PET) Scan.
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
170
Imaging Studies II: Positron Emission Tomography and Scintigraphy
235
Positron Emission Tomography (PET) is a medical imaging technique that provides crucial insights into the body's physiological functions at a molecular level. It is an indispensable resource for diagnosing, staging, and monitoring various illnesses, notably cancer, neurological disorders, and cardiovascular conditions.
Fundamental Principles of PET
Fundamental Principles of PET
235