一种修改的双温度校准方法和辐射度测量装置
Shufang He1, Shuai Li1,2, Caihong Dai1
1Division of Optical Metrology, National Institute of Metrology, Beijing 100029, China.
Materials (Basel, Switzerland)
|July 30, 2025
概括
本研究引入了一种经过修改的双温度校准方法,用于准确测量红外辐射样本的辐射率. 新的方法消除了对参考黑体的需求,简化了这个过程.
科学领域:
- 计量学 计量学 计量学
- 红外热像图 (Infrared Thermography) 是一种红外热像图.
- 材料科学 材料科学 材料科学
背景情况:
- 精确测量红外辐射样本的辐射率至关重要.
- 现有的多温度或两温度校准方法依赖于参考黑体,并假定温度独立的背景辐射,这在实践中往往不是真的.
研究的目的:
- 开发一种经过修改的双温度校准方法和设备,在没有参考黑体的情况下准确测量辐射率.
- 为了克服以前的方法的局限性,背景辐射取决于温度.
主要方法:
- 提出了一种修改后的双温度校准方法,在很短的间隔内设置两个温度点.
- 该方法利用的近似是,在这两个温度下,样品和背景辐射的发射率是相似的.
- 建造了一个实验设施,用于 (8~14) 微米范围的测量.
主要成果:
- 在60°C和85°C的温度下,对0.100,0.500和0.900左右的辐射率的样本进行测量.
- 据报道,不同发射率样本的相对扩展不确定性在85°C时低至1.5%和1.2%.
- 实验结果显示与其他已建立的测量技术一致.
结论:
- 开发的修改后的双温度校准方法对于准确的排放性测量是有效的.
- 这种方法消除了对参考黑体的需求,简化了实际应用.
- 该技术可能适用于温度不敏感的辐射率的样品.
相关概念视频
Temperature Measurement Sites
2.2K
A thermometer measures body temperature. The common sites for measuring body temperature are the oral cavity, axillary region, temporal artery, and skin surface, such as the forehead, abdomen, and axilla. True core body temperature is assessed in the rectum, tympanic membrane, pulmonary artery, esophagus, and urinary bladder.
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
2.2K
Calorimetry
3.2K
When objects at different temperatures are placed in contact with each other but isolated from everything else, they attain thermal equilibrium. A container that prevents heat transfer in or out is called a calorimeter, and the use of a calorimeter to make measurements is called calorimetry. Generally, these measurements involve heat or specific heat capacity. The term "calorimetry problem" is used for any problem where the specified objects are thermally isolated from their...
3.2K
Instrument Calibration
269
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
269
Equipments Used to Measure Body Temperature
1.2K
Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
1.2K
Constant Pressure Calorimetry
86.4K
Calorimetry is a technique used to measure the amount of heat involved in a chemical or physical process or to measure the heat transferred to or from a substance. The heat is exchanged with a calibrated and insulated device called the calorimeter. Calorimetry experiments are based on the assumption that there is no heat exchange between the insulated calorimeter and the external environment. The well-insulated calorimeters prevent the transfer of heat between the calorimeter and its external...
86.4K
Flame Photometry: Lab
363
In a flame photometer, when a solution like potassium chloride is aspirated into the flame, the solvent evaporates, leaving behind dehydrated salt. This salt dissociates into free gaseous atoms in their ground state. Some of these atoms absorb energy from the flame, leading to their excitation. The excited atoms return to the ground state, emitting photons at characteristic wavelengths. Because only electronic transitions are involved, the resulting emission lines are very narrow. The intensity...
363