开发基于嵌入式系统的自动定时提醒装置,以改进自闭柜冷却期管理
Yu Meng1, Ping Gui2, Shunlin Tan1
1Sterile Processing Department, Sichuan Tianfu New Area People's Hospital, Chengdu, China.
PloS one
|August 12, 2025
概括
一个自动定时提醒装置可显著改善冷却时间的遵守,减少无菌加工中的湿包事件. 这种嵌入式系统提高了医疗保健机构的安全性和技术人员满意度.
科学领域:
- 医疗器械开发 医疗器械开发
- 嵌入式系统工程 嵌入式系统工程
- 无菌加工技术的处理技术
背景情况:
- 消毒的物品需要在无菌加工部 (SPD) 进行至少30分钟的冷却期,以防止污染.
- 目前用于管理冷却时间的方法通常是手动的,导致不一致性和潜在风险.
- 减少湿袋和污染对于患者安全至关重要.
研究的目的:
- 开发一个使用嵌入式系统的自动定时提醒装置,用于自闭冷却周期.
- 降低湿包的发病率和相关的污染风险.
- 在无菌加工部 (SPD) 标准化冷却时间管理.
主要方法:
- 使用Arduino编程了一个带有ATMEGA328微控制器,传感器和提醒模块的嵌入式系统设备.
- 该设备安装在灭菌器车上,并使用传统方法对对照组进行了四周的测试.
- 关键指标包括冷却时间的遵守,湿包速率,物品处理频率和技术人员满意度.
主要成果:
- 设备组显示了92.93%的冷却时间合规率,而对照组的72.92%.
- 与对照组 (0.49%) 相比,设备组 (0.18%) 的湿包发病率明显低于对照组 (0.49%).
- 观察到对消毒物品的处理减少,技术人员对设备的满意度提高.
结论:
- 自动定时提醒装置具有成本效益,并受到SPD技术人员的好评.
- 嵌入式系统有效地规范了自闭柜冷却管理,具有广泛的SPD应用潜力.
- 对医疗保健中嵌入式系统的进一步研究可以增强医疗程序.
相关概念视频
Equipments Used to Measure Body Temperature
1.2K
Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
1.2K
Applications of RC Circuits
3.3K
A relaxation oscillator is one of the applications of RC circuits. A neon lamp relaxation oscillator comprises a capacitor, a resistor, a voltage source, and a lamp. The lamp acts like an open circuit, with infinite resistance until the potential difference across the lamp reaches a specific voltage. At that voltage, the lamp acts like a short circuit with zero resistance, and the capacitor discharges through the lamp, thus producing light. Once the capacitor is fully discharged through the...
3.3K
Setting Time of Cement
279
The setting time of cement refers to the process of cement paste transitioning from a plastic state to a solid state. This process is crucial in construction as it dictates the timeframe for concrete placement, compaction, and finishing. The onset of this solidification is termed the initial set, indicating when the paste becomes unworkable. The final set is when the paste has solidified completely, and further handling or manipulation can no longer affect its shape. The cement strength is...
279
Physical Methods for Controlling Microbial Growth: Temperature
247
Heat is a widely used method to control microbial growth by targeting and denaturing cellular proteins, thereby killing or inactivating microbes. This method's effectiveness is quantified using parameters such as the thermal death point (TDP), thermal death time (TDT), and decimal reduction time (D value). TDP represents the lowest temperature at which all microorganisms in a liquid suspension are eliminated within 10 minutes, whereas TDT is the time necessary to achieve sterilization at a...
247
Thermal Stress
2.6K
If the temperature of an object is changed while it is prevented from expanding or contracting, the object is subjected to stress. The stress is compressive if the object expands in the absence of constraint and tensile if it contracts. This stress resulting from temperature change is known as thermal stress. It can be quite large and can cause damage. To avoid this stress, engineers may design components so they can expand and contract freely. For instance, on highways, gaps are deliberately...
2.6K
Assessing Body Temperature - Temporal Artery
667
Here is a stepwise guide to assessing the body temperature at the temporal artery using a temporal artery thermometer
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
667