概括
这项研究测量了Spectralon板块的光谱双向反射分布函数 (BRDF). 99%的斑块在60度发作时表现出最兰伯特式的行为,这对于精确校准颗粒样本至关重要.
科学领域:
- 光学远程传感器是一种远程传感器.
- 材料科学 是一种材料科学.
背景情况:
- 频谱板被广泛用于校准仪器.
- 了解它们的双向反射分布函数 (BRDF) 对于准确的测量至关重要.
研究的目的:
- 测量具有不同反射度 (2-99%) 的 Spectralon 板块的光谱 BRDF.
- 为了评估在不同落入角度 (0°和60°) 的Spectralon斑块的兰伯特行为.
- 评估Spectralon标准对颗粒物样本校准的影响.
主要方法:
- 频谱BRDF测量了从350到2150nm的八个Spectralon板块.
- 测量是在0°和60°的撞击顶角进行的.
- 使用Spectralon标准对颗粒物样本进行校准.
主要成果:
- 40%的Spectralon在正常发病率 (0°) 中表现出最多的兰伯特行为.
- 99%的Spectralon在60度的入射角下表现出最好的兰伯特式行为.
- 颗粒样本的校准反射率值 (REFF) 在使用99%的Spectralon斑块时是最高的.
结论:
- 选择Spectralon板块会影响颗粒样本的绝对反射度校准.
- 建议使用99%的Spectralon板来达到更高的校准反射值,特别是在非正常发生条件下.
相关概念视频
Raman Spectroscopy: Overview
600
The underlying principle of Raman spectroscopy is based on the interaction between light and matter, specifically molecules' inelastic scattering of photons. When a monochromatic beam of light, typically from a laser source, interacts with a sample, most scattered light has the same frequency as the incident light. This is known as Rayleigh scattering.
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
600
Raman Spectroscopy Instrumentation: Overview
532
A conventional Raman spectrophotometer includes a laser source, a sample holding system, a wavelength selector, and a detector.
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
532
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
555
Attenuated total reflectance (ATR) infrared spectroscopy is a powerful analytical technique used to study the composition of materials. It is widely employed in chemistry, materials science, forensic science, and other fields where sample characterization is required. ATR has several advantages over traditional transmission IR spectroscopy, including the requirement of little to no sample preparation and the ability to analyze a wide range of samples.
The ATR process begins by directing a beam...
The ATR process begins by directing a beam...
555