Jove
Visualize
联系我们
JoVE
x logofacebook logolinkedin logoyoutube logo
关于 JoVE
概览领导团队博客JoVE 帮助中心
作者
出版流程编辑委员会范围与政策同行评审常见问题投稿
图书馆员
用户评价订阅访问资源图书馆顾问委员会常见问题
研究
JoVE JournalMethods CollectionsJoVE Encyclopedia of Experiments存档
教育
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab Manual教师资源中心教师网站
使用条款与条件
隐私政策
政策

相关概念视频

Quantifying Work02:30

Quantifying Work

As a system undergoes a change, its internal energy can change, and energy can be transferred from the system to the surroundings, or from the surroundings to the system.
Distributed Loads01:19

Distributed Loads

Distributed loads are a common type of load that engineers and scientists encounter in various practical situations. Distributed loads often refer to a type of load spread over a surface or a structure and can be modeled as continuous force per unit area.
For example, consider a bookshelf filled with books stacked vertically adjacent to each other. The weight of the books is evenly distributed over the length of the shelf. As a result, the pressure at different locations on the surface of the...
Distribution Reliability and Automation01:25

Distribution Reliability and Automation

Distribution reliability in electrical power systems is critical for ensuring an uninterrupted power supply to consumers at minimal cost. According to IEEE Standard Terms, reliability is the probability that a device will function without failure over a specified time period or amount of usage. For electric power distribution, this translates to maintaining continuous power supply and addressing customer concerns over power outages. Several indices, as defined by IEEE Standard 1366-2012, are...
Flow Table Test01:12

Flow Table Test

The flow table test is an established method used to assess the workability of concrete, particularly useful for evaluating highly flowable concrete mixes. This test employs an apparatus that consists of a wooden board topped with a steel plate, collectively weighing 35 pounds. The board is connected to a base via a hinge and measures 27.6 inches on each side.
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
Run Charts01:12

Run Charts

Run charts serve as an essential instrument for visualizing the performance of various processes over time, enabling the identification of trends and patterns crucial for quality improvement. These charts map out a series of data points chronologically, offering insights into the stability and efficiency of a process. A run chart's creation involves plotting data points on a graph, with the time intervals on the horizontal axis and the specific measurements on the vertical axis. For example,...
Scale-Up Processes01:14

Scale-Up Processes

The scale-up of microbial fermentation processes is essential in industrial biotechnology, allowing the transition from laboratory-scale experiments to commercial-scale production while aiming to maintain product yield and quality. This process requires meticulous adjustment of equipment design, process parameters, and contamination control strategies to accommodate increasing culture volumes.At the laboratory scale, cultures are typically maintained in 1 to 10-liter glass or autoclavable...

您也可能阅读

相关文章

通过共同作者、期刊和引用图与本文相关的文章。

排序
Same author

Clinical validation of a high-definition mid-position magnetic resonance imaging approach for lung radiotherapy planning.

Medical physics·2026
Same author

Motion mitigation in positron-emission tomography guided radiotherapy delivered on a magnetic resonance imaging-linear accelerator.

Physics and imaging in radiation oncology·2026
Same author

MRI-guided radiotherapy: is the best still to come?

Physics in medicine and biology·2026
Same author

The first experimental implementation of real-time dose-guided MLC tracking using an MRI-linac.

Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology·2026
Same author

X-ray beam quality checks of a magnetic resonance-guided linear accelerator with its on-board megavoltage imager.

Physics and imaging in radiation oncology·2026
Same author

Real-time dose reconstruction and dose coverage forecasting using the magnetic resonance linear accelerator.

Physics and imaging in radiation oncology·2026

相关实验视频

Updated: May 10, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
08:17

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy

Published on: June 7, 2015

15.8K

开发和演示端到端测试的分数内运动管理的工作流程

Stijn Oolbekkink1, Jochem W H Wolthaus1, Bram van Asselen1

  • 1Department of Radiotherapy, University Medical Center Utrecht, Utrecht, the Netherlands.

Medical physics
|August 24, 2025
PubMed
概括

使用运动包含剂量分布开发了用于端到端测试分数内运动控制的工作流程的新方法. 这些新方法,APriMI和PostMI,可以通过实时运动补偿来准确评估治疗的准确性.

科学领域:

  • 医学物理
  • 辐射瘤学
  • 图像指导治疗

背景情况:

  • 在MR-linacs上使用分量内运动管理技术 (光束隔离,分量内偏移校正) 来减少运动的剂量影响.
  • 剩余运动可能导致规划剂量的偏差,常规端到端 (E2E) 测试不考虑这种已知的运动.
  • 这限制了动作管理工作流程的准确性评估.

研究的目的:

  • 开发和展示用于E2E测试的参考剂量分布中已知运动的新方法.
  • 为了能够准确的E2E测试分数内运动管理的工作流程.

主要方法:

  • 开发了包括运动在内的两个参考剂量分布:包括运动在内的先验 (APriMI) 和包括运动在内的后期 (PostMI).
  • APriMI使用已知的周期性运动来估计和结合运动对剂量的影响;PostMI使用外部触发器来与非周期性或漂移运动的运动相关联.
  • 在使用膜剂量计的Unity MR-linac上评估静态,无控制和运动控制 (Gated,IDC) 场景,将测量剂量与静态,APriMI和PostMI参考值进行比较.

主要成果:

  • 在静态基准场景中,测量剂量与参考剂量之间存在很高的一致性.
  • 在使用APriMI或PostMI (几何偏移<0. 3mm,玛指数>95. 5%) 的运动包含场景中,测量剂量和计算剂量之间观察到很好的一致性.
关键词:
核磁共振指导放射治疗端到端测试在分数内进行运动管理

更多相关视频

In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
07:43

In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy

Published on: July 2, 2021

3.1K
Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
10:23

Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management

Published on: June 23, 2023

3.0K

相关实验视频

Last Updated: May 10, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
08:17

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy

Published on: June 7, 2015

15.8K
In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
07:43

In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy

Published on: July 2, 2021

3.1K
Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
10:23

Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management

Published on: June 23, 2023

3.0K
  • 使用静态剂量分布作为参考结果导致显著的偏差 (几何偏移高达9. 0毫米,马传递率低至17. 8%).
  • 结论:

    • APriMI和PostMI的剂量分布使得E2E可以有效地测试分数内运动控制的工作流程.
    • 与静态参考不同的是,运动包含的参考剂量分布与测量结果有很强的一致性.
    • 准确评估E2E的准确性需要在治疗过程中包含实际运动的参考剂量文件.