在高功率短时间切除过程中,柔性尖端与非柔性尖端导管之间的微栓塞信号差异
Yoshiaki Mizutani1, Satoshi Yanagisawa2, Mizuki Ichikawa3
1Department of Cardiology, Yokkaichi Municipal Hospital, Yokkaichi, Japan.
Journal of cardiovascular electrophysiology
|August 26, 2025
概括
与非柔性尖端导管相比,这种新型的柔性尖端导管在高功率短时间 (HPSD) 移除过程中显著降低了微栓塞信号. 这表明柔性尖端导管可能为心房动提供更安全的HPSD切除.
科学领域:
- 心血管干预
- 医疗设备技术
- 电生理学
背景情况:
- 一种新型的柔性尖端除菌导管具有独特的远端, 旨在优化灌流量.
- 高强度短时间 (HPSD) 切除是对心房的有效治疗方法.
- 评估微栓塞信号 (MES) 对于评估除手术的安全性至关重要.
研究的目的:
- 在柔性尖端和非柔性尖端除导管之间比较HPSD射频除过程中MES的发生率.
- 评估新型柔性尖端导管在减少栓塞事件中的安全性和有效性.
主要方法:
- 对70名首次接受HPSD切除 (50W) 的患者进行了回顾性分析.
- 在肺静脉隔离过程中使用带超声波多普勒对MES的临床评估.
- 用两种类型的导管对猪心脏组织进行实验性切除以评估MES.
主要成果:
- 与非柔性尖端组相比,柔性尖端组的MES,MES每点和MES每秒的数量显著降低 (p < 0. 001).
- 在倾向性得分匹配后,MES在柔性尖端组 (p = 0.013) 中仍然明显较低,特别是泡型MES (p = 0.023).
- 实验研究证实,使用柔尖导管的MES显著减少 (p = 0.002).
结论:
- 与50W的非柔性尖端导管相比,柔性尖端导管的切除显著减少了MES检测.
- 这些发现表明,柔性尖端导管可以实现更安全的HPSD切除手术.
- 灵活尖端导管的优化灌设计可能有助于减少栓塞事件.
更多相关视频
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
13.4K
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
14.9K
相关概念视频
π Electron Effects on Chemical Shift: Overview
1.5K
An applied magnetic field causes loosely bound π-electrons in organic molecules to circulate, producing a local or induced diamagnetic field over a large spatial volume. As the molecules tumble in solution, the field generated by π-electrons in spherical substituents results in a zero net field. However, the net field generated by π-electrons in non-spherical substituents is not zero. The effect of this induced field depends on the orientation of the molecule with respect to B0,...
1.5K
Double Resonance Techniques: Overview
870
Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
Spin decoupling is usually achieved by...
870
Instrumentation Amplifier
1.3K
An electrocardiography (ECG) machine is an essential piece of medical equipment used to monitor the electrical activity of the heart. It operates by detecting small electrical changes on the skin that result from the depolarization of the heart muscle during each heartbeat. However, these signals are in the microvolt range and can be easily overwhelmed by noise or interference.
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
1.3K
Clipper Circuit
1.0K
A clipper circuit is a fundamental wave-shaping device that harnesses the unique properties of diodes to alter and control waveform characteristics. This technology is widely used in electronic devices, especially in television and radar communication systems, where it enhances waveform modulation in both transmitters and receivers.
The operation of a clipper circuit can be exemplified by analyzing a dual-clipper configuration setup that integrates two ideal diodes, each paired with a biasing...
The operation of a clipper circuit can be exemplified by analyzing a dual-clipper configuration setup that integrates two ideal diodes, each paired with a biasing...
1.0K
Cut-off Frequency of BJT
1.5K
Cut-off frequencies in Bipolar Junction Transistors (BJTs) mark the transition between the signal's pass band and stop band, influencing their performance in amplifying or attenuating frequencies. These frequencies are crucial for designing BJTs to meet specific operational requirements in electronic circuits.
Alpha Cut-Off Frequency: Pertinent to the common-base configuration, the alpha cut-off frequency defines the upper-frequency limit at which the current gain, alpha, remains stable. As...
Alpha Cut-Off Frequency: Pertinent to the common-base configuration, the alpha cut-off frequency defines the upper-frequency limit at which the current gain, alpha, remains stable. As...
1.5K
Biasing of FET
1.0K
Biasing a Junction Field Effect Transistor (JFET) is crucial for setting operational parameters and ensuring efficient functioning in electronic circuits. JFETs are characterized by using a single carrier type in N-channel or P-channel configurations, where the channel is surrounded by PN junctions. These junctions are central to the device's ability to control current flow.
In an N-channel JFET, the structure consists of N-type material forming the channel on a P-type substrate, with the...
In an N-channel JFET, the structure consists of N-type material forming the channel on a P-type substrate, with the...
1.0K