在恒定负载循环运动中,CORE可穿戴传感器的有效性
Brayden C McLaughlin1, Joseph T Aguilera1, Andrew C D'Lugos2
1Department of Kinesiology, California State University, Chico, Chico, CA, USA.
Journal of thermal biology
|August 28, 2025
概括
这种可穿戴设备在运动中高估了核心体温 (TC). 皮下脂肪厚度影响了设备的准确性,表明TC估计的潜在偏差.
科学领域:
- 运动生理学
- 生物医学工程
- 环境健康
背景情况:
- 精确的核心体温 (TC) 监测对于在炎热条件下进行体力活动时预防与热有关的疾病至关重要.
- 可穿戴技术提供了一个非侵入性的方法来估计TC,但它的有效性需要彻底调查.
研究的目的:
- 评估CORE可穿戴传感器在热环境中运动时估计核心体温 (TC) 的有效性.
- 调查皮下脂肪厚度在CORE传感器TC测量中观察到的潜在偏差中的作用.
主要方法:
- 24名健康参与者 (13名男性,11名女性) 在炎热的环境中以60%的最大有氧功率进行60分钟的自行车运动.
- 使用CORE传感器测量核心体温 (TC) 和皮肤温度 (TSK),并与直肠和皮肤表面探针进行比较.
- 使用超声波测量皮下脂肪厚度,以评估其与TC偏差的相关性.
主要成果:
- CORE 传感器显示TC的总体估值不显著,但在运动后半段显著高估.
- 在整个运动期间,该设备还显著高估了皮肤温度 (TSK).
- 在运动的第三和第四季度中,米达克西拉皮下脂肪厚度与TC高估偏差呈正相关性.
结论:
- CORE 传感器显示,与长时间在炎热环境下进行的直肠测量相比,核心体温 (TC) 的估计过高.
- 皮肤温度升高 (TSK) 和局部皮下脂肪厚度似乎是导致CORE传感器TC估计偏差的潜在因素.
更多相关视频
08:22Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
10.7K
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
669
相关概念视频
Exercise and Cardiac Output
1.3K
Regular physical activity is essential for maintaining cardiovascular health, with aerobic exercises being particularly effective. According to the American Heart Association, 150 minutes of moderate to intense aerobic exercise per week is recommended for a healthy heart. Aerobic activities may include brisk walking, running, bicycling, cross-country skiing, and swimming, ideally performed three to five times per week.
Sustained exercise increases the muscles' oxygen demand, which can be...
Sustained exercise increases the muscles' oxygen demand, which can be...
1.3K
Exercise Stress Test
489
Introduction
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
489
Equipments Used to Measure Body Temperature
1.2K
Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
1.2K
Exercise and Cardiovascular Response
1.1K
Exercise significantly impacts cardiovascular response, which is crucial for understanding patient health and designing effective treatment plans.
Light to moderate physical activity initiates a series of interconnected responses in the body. The heart rate modestly increases in anticipation of the workout, followed by widespread vasodilation as oxygen consumption by skeletal muscles increases. This results in decreased peripheral resistance, increased capillary blood flow, and accelerated...
Light to moderate physical activity initiates a series of interconnected responses in the body. The heart rate modestly increases in anticipation of the workout, followed by widespread vasodilation as oxygen consumption by skeletal muscles increases. This results in decreased peripheral resistance, increased capillary blood flow, and accelerated...
1.1K
Thermal Stress
2.6K
If the temperature of an object is changed while it is prevented from expanding or contracting, the object is subjected to stress. The stress is compressive if the object expands in the absence of constraint and tensile if it contracts. This stress resulting from temperature change is known as thermal stress. It can be quite large and can cause damage. To avoid this stress, engineers may design components so they can expand and contract freely. For instance, on highways, gaps are deliberately...
2.6K
Assessing Body Temperature - Axilla
661
Procedural Guide for Assessing Axillary Body Temperature using a Digital Thermometer:
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
661