用ICP-MS和pXRF进行横截面研究
Michael Anastario1, Delaney Watkins2, Aaron J Specht3
1Department of Health Sciences, Northern Arizona University, Flagstaff, AZ, USA. Michael.anastario@nau.edu.
BMC research notes
|September 2, 2025
概括
这项研究使用ICP-MS量化了哈斯梨部分的金属度,并评估了便携式X射线光 (pXRF) 用于元素分析. pXRF显示了快速,现场选梨金属含量的潜力.
科学领域:
- 农业科学
- 环境科学
- 分析化学
背景情况:
- 农产品中的金属度是非常不同的.
- 了解元素成分对于食品安全和质量至关重要.
研究的目的:
- 量化哈斯梨中和外中的金属度.
- 评估便携式X射线光 (pXRF) 作为元素分析的快速查工具.
主要方法:
- 诱导合等离子体质谱 (ICP-MS) 用于精确的金属量化.
- 便携式X射线光器 (pXRF) 用于快速的元素查.
- 根据原产地和农业实践分层分析了24个梨样本.
主要成果:
- ICP-MS检测出多种金属 (K,Mg,Ca,Fe,Co,Ni,Zn,Hg) 在外皮中的度较高.
- 根据来源和养殖方法而异的确定元素的同时存在模式 (例如,Co-Ni,Mg-Fe-Mo-Hg).
- 对于K,Ca,Fe和Zn,pXRF与ICP-MS呈现中等相关性,这表明其具有实地查的潜力.
结论:
- 提供梨的基线金属度数据.
- 证明了pXRF在农业研究中的快速实地元素分析的可行性.
- 突出了元素分析对环境监测和农业研究的价值.
更多相关视频
08:43PTR-ToF-MS Coupled with an Automated Sampling System and Tailored Data Analysis for Food Studies: Bioprocess Monitoring, Screening and Nose-space Analysis
Published on: May 11, 2017
12.5K
06:53Author Spotlight: Technologies and Challenges in Elemental Analysis of Food Samples
Published on: December 22, 2023
2.5K
相关概念视频
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
931
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
931
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
617
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
617