"低级红光"近视控制装置的放射测量和安全性评估
Josh Richards1, Jennifer J Hunter2, Javier Gantes-Nuñez3
1School of Optometry, Indiana University, Bloomington, Indiana, USA.
概括
控制近视的红光治疗可能会对视网膜造成危害. 这项研究发现近视控制装置
科学领域:
- 眼科 眼科
- 生物医学光学
- 医疗器械
背景情况:
- 低水平的红光疗法可以减缓儿童的轴延长.
- 对于一些用于近视控制的红灯装置的安全性存在担忧.
- 了解光学特征和潜在的危害对于设备安全至关重要.
研究的目的:
- 为了研究特定的红光近视控制装置 (SECONEE sky-n1201a) 的光束特性.
- 评估设备的光辐射对视网膜的潜在危害.
- 将视网膜暴露水平与已确定的安全标准进行比较.
主要方法:
- 通过人工瞳孔 (1-9毫米) 量化的光束分歧和总流量.
- 根据瞳孔间距离设置测量出口度.
- 视网膜暴露水平与美国国家标准协会 (ANSI) Z80.36-2021限制值进行比较.
主要成果:
- 该装置发射了具有可变出口度的共聚红光束 (655 nm).
- 随着人造瞳孔直径的增加,
- 视网膜暴露超过ANSI热和光化学限制,分别为瞳孔直径>1mm和>5mm.
- 对于大于5毫米的瞳孔,在100秒内达到潜在的光化学危险.
结论:
- 该设备的光学设计导致视网膜暴露取决于瞳孔大小,适应和眼睛运动.
- 视网膜辐射暴露可能超过ANSI安全限值,挑战"低水平"的指定.
- 需要对红灯近视控制装置进行进一步的安全评估.
相关概念视频
Biological Effects of Radiation
15.5K
All radioactive nuclides emit high-energy particles or electromagnetic waves. When this radiation encounters living cells, it can cause heating, break chemical bonds, or ionize molecules. The most serious biological damage results when these radioactive emissions fragment or ionize molecules. For example, α and β particles emitted from nuclear decay reactions possess much higher energies than ordinary chemical bond energies. When these particles strike and penetrate matter, they...
15.5K
Radiological Investigation III: Pulmonary Angiogram and PET Scan
732
Radiological investigations are paramount in the diagnosis and management of various pulmonary diseases. Two essential investigations are the Pulmonary Angiogram and the Positron Emission Tomography (PET) Scan.
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
732