扩大单颗粒ICP-MS的线性动态范围用于微塑料的量化
George C Caceres1,2, Monique E Johnson1, John L Molloy1
1Chemical Sciences Division, Material Measurement Laboratory, National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, Maryland 20899-1070, United States.
Analytical chemistry
|September 5, 2025
概括
单颗粒感应合等离子体质谱 (spICP- MS) 已被优化为准确的微塑性分析. 这种方法可以更好地检测和量化高达5μm的聚乙烯微塑料,从而提高环境监测能力.
科学领域:
- 环境科学
- 分析化学
- 材料科学
背景情况:
- 微塑料的积累对人类健康构成风险,需要先进的分析方法来检测和描述.
- 由于其检测能力和速度,单颗粒感应合等离子体质谱 (spICP-MS) 对微塑料分析具有前景.
- 由于碳含量限制和运输效率低下,目前的spICP-MS方法难以准确地测量和量化微塑料,特别是大于3μm的微塑料.
研究的目的:
- 将spICP-MS的线性动态范围扩展到5μm以准确量化聚乙烯微塑料 (PS MP).
- 通过spICP-MS提高微塑料颗粒数量 (PNC) 的检测和量化.
- 通过优化样品引入和操作参数来解决spICP-MS分析中较大的微塑料颗粒的局限性.
主要方法:
- 使用spICP-MS的高效样本引入系统.
- 优化了13C信号并减少了雾化器气体流量以提高更大的颗粒的运输.
- 通过监测其碳含量 (13C+) 来定量的聚乙烯微粒 (PS MPs).
主要成果:
- 在2至5μm的PS MP中,可靠量化颗粒数度 (PNC) 在库存度的20%以内.
- 在混合物中的PS MP的准确量化,不论PNC比率如何.
- 将spICP-MS的精确量化范围扩展到更大的颗粒大小 (高达5μm).
结论:
- 优化spICP-MS参数,包括降低雾化器气流,显著改善较大的微塑料的分析.
- 增强的spICP-MS方法可提供微塑料颗粒数量的可靠量化,这对于环境和健康风险评估至关重要.
- 这一进步为解决与微塑料污染相关的问题提供了更强大的分析工具.
相关概念视频
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
928
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
928
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
614
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
614