相关实验视频
Updated: Jan 18, 2026

05:22
Iris Fixation via External Pentagram Suturing
Published on: May 5, 2022
1.8K
雅马内侧翼脑内固定透镜中的触觉-光学骨折:一个案例系列
Rina Su1,2, Paras P Shah1,2, Elmira Baghdasaryan1,2
1Department of Ophthalmology, Northwell Health Eye Institute, Great Neck, NY, USA.
Journal of ophthalmic & vision research
|September 10, 2025
概括
在Yamane侧边肌内触觉固定手术 (FIHFS) 期间CT Lucia 602眼内透镜 (IOL) 的结构性弱点导致光学断裂和微骨折. 建议对这种手术技术进行仔细的内腔镜材料选择.
科学领域:
- 眼科医生 眼科 眼科
- 医疗设备工程 医疗设备工程
- 手术技巧 手术技巧
背景情况:
- 雅马内侧边肌内触觉固定手术 (FIHFS) 是用于眼内透镜 (IOL) 植入的公认技术.
- 内腔镜的结构完整性对于成功的手术结果和患者安全至关重要.
研究的目的:
- 在CT Lucia 602 IOL的触觉光学结处报告结构性弱点的情况.
- 为了记录在Yamane FIHFS期间发生的光学断裂和微骨折.
主要方法:
- 一个单一的外科医生对五名接受了Yamane FIHFS的患者进行了回顾性图表审查.
- 数据收集包括患者人口统计,手术技术,术后结果和12个月的随访期间的不良事件.
主要成果:
- 在CT Lucia 602 IOL的触觉-光学结处发现了5例结构性弱点.
- 两个病例涉及光学断裂 (一个需要内腔镜更换,一个需要扩张),三个病例显示光学微骨折.
- 所有的骨折都发生在领先触觉的结合处.
结论:
- 在触觉-光学结处的光学断裂和微骨折是与Yamane技术相关的不常见并发症.
- 这种并发症以前没有在CT Lucia 602 IOL中报告过.
- 建议仔细选择IOL材料,以减轻这种手术方法的潜在并发症.
相关概念视频
Glaucoma: Overview
1.3K
Glaucoma is an eye condition characterized by increased intraocular pressure that damages the retina and optic nerve, leading to irreversible blindness if left untreated. The human eye has various components, including the cornea, iris, pupil, lens, and optic nerve. Aqueous humor is secreted by the epithelium of the ciliary body in the posterior chamber and flows through the trabecular meshwork and canal of Schlemm, maintaining normal intraocular pressure. The trabecular meshwork and the canal...
1.3K
Angle Closure Glaucoma: Treatment
1.2K
Angle-closure glaucoma, or closed-angle glaucoma, is an eye condition where the iris bulges out and blocks the iridocorneal angle, resulting in a buildup of aqueous humor and increased intraocular pressure. Immediate medical attention is necessary due to the sudden onset of symptoms. The treatment for angle-closure glaucoma includes short-term and long-term approaches. Short-term treatment involves using eye drops like pilocarpine to lower intraocular pressure by increasing aqueous humor...
1.2K
Open Angle Glaucoma: Treatment
963
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...
963
Focusing of Light in the Eye
5.4K
Light rays enter the eye through the cornea, a transparent dome-shaped tissue that is the eye's outermost layer. The cornea bends or refracts, light rays traveling to the pupil. The shape of the cornea determines how much of the light is bent and whether the image will be focused correctly on the retina at the back of the eye. Once the light has passed through both refraction layers, it converges into a single focal point onto a small area. This is where photoreceptors start transforming...
5.4K