在实验室中,BiVACOR全人工心脏在连续和脉冲性流动中的血液兼容性
Antony P McNamee1, Daniel Timms2, Frank Nestler2
1Biorheology Research Laboratory, Griffith University, Gold Coast, Australia.
Artificial organs
|September 17, 2025
概括
在体外,BiVACOR的整体人工心脏 (TAH) 与参考具有可比的血液相容性. 这种新设备有效地支持血液循环,而不会显著影响血液参数,无论流动模式如何.
科学领域:
- 生物医学工程 生物医学工程
- 心血管研究研究心血管研究
- 医疗器械 医疗器械
背景情况:
- 心脏移植受限于供体器官的可用性.
- 机械循环支持装置是晚期心力衰竭的关键替代品.
- BiVACOR的全人工心脏 (TAH) 使用一种新型离心旋转器设计,可以提供全心脏替代.
研究的目的:
- 为了评估BiVACOR的TAH的体外血红相容性.
- 为了比较连续流 (CF) 与脉动流 (PF) 模式中的血液相容性.
- 为了评估血液兼容性与临床批准的参考相比.
主要方法:
- 牛血在体外循环中以5L/分钟和100mmHg的速度在6小时内循环.
- 在CF和PF模式下测试TAH,并与CentriMag参考进行比较.
- 血液分析包括血液学,血自由血红蛋白和·维勒布兰德因子 (vWF) 多度量.
主要成果:
- 在TAH (CF和PF) 和参考之间,正常化血液溶解指数是可比的.
- 基本血液学和vWF多元仪显示线性变化,但在流动模式或设备之间没有显著的变化.
- 血液溶解水平较低,表明TAH具有良好的血液相容性.
结论:
- 在体外,BiVACOR的TAH显示出与临床批准的参考相比的血液兼容性.
- 该设备有效地支持全身和肺循环.
- 脉动性流动不会对血红相容性产生负面影响,这表明它适合未来的应用.
更多相关视频
09:41Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
10.0K
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
7.8K
相关概念视频
Blood Flow
Blood is pumped by the heart into the aorta, the largest artery in the body, and then into increasingly smaller arteries, arterioles, and capillaries. The velocity of blood flow decreases with increased cross-sectional blood vessel area. As blood returns to the heart through venules and veins, its velocity increases. The movement of blood is encouraged by smooth muscle in the vessel walls, the movement of skeletal muscle surrounding the vessels, and one-way valves that prevent backflow.
Equipments Used To Measure Blood Pressure
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Autoregulation of Blood Flow
Autoregulation mechanisms are characterized by their inherent capacity for self-regulation without necessitating specific nervous stimulation or endocrine control. These mechanisms facilitate the adjustment of blood flow and, therefore, perfusion specific to each tissue region. This self-regulation encompasses chemical signals and myogenic controls.
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation.
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation.