确定在沙特阿拉伯使用的葡萄糖计的精度
Shoug A Al-Othman1, Zahra H Al-Zaidany1, Shahad H Al-Ghannam1
1College of Medicine, Imam Abdulrahman Bin Faisal University, Dammam 31441, Eastern Province, Saudi Arabia.
Sensors (Basel, Switzerland)
|September 19, 2025
概括
在经过测试的产品中,Accu-Chek Instant 是唯一符合ISO准确性标准的血糖计. 维生素C显著干扰葡萄糖读数,影响准确度.
科学领域:
- 临床化学 临床化学
- 医疗器械技术 医疗器械技术
背景情况:
- 血糖计的可靠性至关重要,但市场上准确性和精度的后期测试往往缺乏.
- 确保血糖监测设备的一致性能对于患者护理至关重要.
研究的目的:
- 为了评估沙特阿拉伯常用的血糖计的精度.
- 为了研究维生素C,乙氨基和麦芽糖对葡萄糖读数的干扰作用.
主要方法:
- 对Accu-Chek即时,值班Sharp和ConTour血糖计进行比较分析研究.
- 从健康志愿者的血液样本中分析了正常,高和低血糖水平.
- 干扰测试涉及多种度的维生素C,乙氨基和麦芽糖的样品.
主要成果:
- 在所有测试的葡萄糖度 (ICC 0.95-1, p < 0.001) 上,Accu-Chek Instant 满足了国际标准化组织 (ISO) 的精度标准.
- 维生素C对葡萄糖读数的干扰比乙氨基和麦芽糖更显著.
- 其他测试的血糖计并不始终符合ISO精度标准.
结论:
- 葡萄糖计的精度和干扰效应对于预防胰岛素注射错误至关重要.
- 准确的葡萄糖读数对于管理糖尿病和避免严重并发症至关重要.
相关概念视频
Uncertainty in Measurement: Accuracy and Precision
99.9K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.
99.9K
Uncertainty in Measurement: Reading Instruments
50.3K
Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
50.3K
Accuracy and Precision
14.0K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value. Highly accurate...
14.0K
Errors occurring during blood pressure monitoring
1.3K
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
1.3K
Data Validation
578
Method validation is a crucial process in analytical chemistry designed to confirm that a given method consistently produces reliable and high-quality results. This process is essential when a method is applied to different sample matrices or when procedural modifications are made, ensuring that the results meet acceptable standards across various applications.
Key parameters for method validation include:
Key parameters for method validation include:
578
Assessment of blood pressure in brachial artery(one-step method)
1.1K
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
1.1K