使用FlexNIRS和机器学习在 carotid endarterectomy期间进行非侵入性连续血压预测
Zahra Einalou1, Mehrdad Dadgostar1, Kuan-Cheng Wu1
1Massachusetts General Hospital, Harvard Medical School, Athinoula A. Martinos Center for Biomedical Imaging, Boston, Massachusetts, United States.
近红外光谱光斑显微镜 (NIRS-PPG) 提供了一种用于连续血压监测的非侵入性方法. 这项研究表明,使用机器学习的NIRS-PPG可以准确地估计平均动脉压,并有可能取代动脉直线.
科学领域:
- 生物医学工程 生物医学工程
- 医疗器械 医疗器械
- 生理监测 生理监测
背景情况:
- 持续监测血压对于患者的稳定至关重要.
- 近红外光谱光斑显微镜 (NIRS-PPG) 是传统动脉线 (A线) 的一种非侵入性的替代方案.
研究的目的:
- 评估可穿戴NIRS-PPG设备 (FlexNIRS) 的有效性,以估计平均动脉压 (MAP).
- 使用NIRS-PPG信号对MAP估计进行线性和高斯过程回归 (GPR) 模型进行比较.
主要方法:
- 从10名患者双边收集NIRS-PPG信号,这些信号是在心动脉内关节切除术期间采集的.
- 开发了特定对象的线性回归和GPR模型,使用NIRS-PPG特征和心率来预测MAP.
- 动脉线读数被用作黄金标准进行比较.
主要成果:
- 所有模型都表现出强的性能,R平方值>= 0.75.
- 一个两种特征的GPR模型实现了0.78的R平方,进一步改进到0.82的第三个特征.
- 在准确度指标 (R平方,MAE,RMSE) 中观察到有统计学意义的改善.
结论:
- 这项初步研究证实了使用NIRS-PPG和机器学习进行非侵入性MAP估计的可行性.
- 在重症监护机构中,NIRS-PPG方法显示为血压监测后A-line移除的切实可行的替代方案.
更多相关视频
07:25Predicting Amputation using Local Circulating Mononuclear Progenitor Cells in Angioplasty-treated Patients with Critical Limb Ischemia
Published on: September 22, 2020
05:51Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
相关概念视频
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.