无线仪器内 (WalkinSense系统) 的有效性用于测量步态指标
Melanie Eckelt1, Jennifer Fayad1, Anne Backes1
1Department of Precision Health Luxembourg Institute of Health Strassen Luxembourg.
Journal of experimental orthopaedics
|September 29, 2025
概括
步行感应内准确地测量了步行时空变量,显示了良好的到极好的一致性. 然而,如果没有显著的改进,它不适合动力学分析.
科学领域:
- 生物力学 生物力学
- 康复工程 康复工程
- 可穿戴技术可穿戴技术
背景情况:
- 步行分析对于评估运动障碍和骨科康复治疗疗效至关重要.
- 可穿戴式传感器为家庭步态监测提供了潜力,但需要验证.
研究的目的:
- 为了验证无线仪表内 (WalkinSense) 的准确性和可靠性,用于步态分析.
主要方法:
- 104名健康参与者在各种速度和坡度进行了步行和跑步试验.
- 使用WalkinSense系统,仪器跑步机和3D运动捕捉系统收集了步行数据.
- 统计分析包括布兰德-阿尔特曼分析,MAPE和ICC,以评估协议和一致性.
主要成果:
- 在行走过程中,WalkinSense对时空变量表现出良好的一致性 (ICC > 0.88,MAPE < 5%).
- 在运行过程中,对于时空变量发现了可接受的良好协议.
- 在所有条件下,所有动力学变量都观察到差的一致性.
结论:
- 步行感应系统是一个有效的工具,用于量化步行时的时空步行参数.
- 该系统需要进一步开发以进行准确的动力学测量.
相关概念视频
Equipments Used To Measure Blood Pressure
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Electronic Distance Measuring Instruments
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over short distances...