在过渡状态下心率测量的准确性:对可穿戴设备进行现实生活监测的验证研究
Catharina Nina Van Oost1, Federica Masci1, Adelien Malisse1
1Department of Biosystems, KU Leuven, 3000 Leuven, Belgium.
Sensors (Basel, Switzerland)
|October 29, 2025
概括
对于心率监测的可穿戴设备的准确性各不相同. 胸带表现最好,而一些手腕设备如Fitbit和Garmin在动态条件下提供适度的准确性,但在快速变化时扎.
科学领域:
- 生物医学工程 生物医学工程
- 可穿戴技术可穿戴技术
- 生理监测 生理监测
背景情况:
- 可穿戴设备非常受欢迎,用于健康和压力监测.
- 对于这些设备在现实世界动态条件下的准确性存在担忧.
研究的目的:
- 为了验证各种可穿戴设备的心率测量精度与12导电心电图相比.
- 在模拟现实生活动态活动和暂时心率变化期间评估设备性能.
主要方法:
- 对比一个胸部佩戴的 (Zephyr BioHarness 3.0) 和七个手腕佩戴的设备 (EmbracePlus,Fitbit Charge 5,Fitbit Sense 2,Garmin Vivosmart 4,WHOOP 4.0,Withings Scanwatch) 与一个12导电心电图.
- 使用20分钟的休息和不同强度的步行协议来模拟动态条件.
- 在不同的分辨率 (1s,10s,60s) 和过渡状态期间评估准确性.
主要成果:
- 胸部佩戴的Zephyr设备在所有动态条件下表现出强的性能.
- 在手腕设备中,Fitbit Charge 5和Sense 2的整体准确度是最高的.
- 戴在手腕上的设备,特别是在短暂状态 (运动,快速HR变化) 中,显示精度降低,尽管更大的平均窗口改善了结果.
结论:
- 腕式可穿戴设备对于平均/趋势心率监测比急性动态更好.
- 适合Fitbit和Garmin设备,在动态条件下提供适合的中等精度.
- 在有效的健康和压力监测应用中,特定环境的验证和明智的设备选择至关重要.
相关概念视频
Pulse rhythm
1.3K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.3K
Holter Monitor: 24-Hour Monitoring
2.0K
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
2.0K
Special considerations while measuring pulse
895
Assessing a patient's pulse is a fundamental skill in healthcare, but certain situations require special attention:
895
Factors Influencing Heart Rate
6.1K
The heart rate, or pulse rate, is a vital indicator of cardiovascular health. It reflects the number of times the heart beats per minute. Various physiological and environmental factors influence heart rate, increasing or decreasing cardiac output. Understanding these factors is crucial for assessing heart function and identifying potential health issues.
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
6.1K
Errors occurring during blood pressure monitoring
1.3K
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
1.3K
Equipments Used To Measure Blood Pressure
3.0K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.0K