装有电容型压力映射传感器板的闭塞力测量装置的内部和内部检查员可靠性
Yoshihiro Kugimiya1, Masaharu Murakami1, Yu Nakano1
1Department of Dentistry and Oral Surgery, National Center for Geriatrics and Gerontology, Obu, Japan.
Journal of prosthodontic research
|November 3, 2025
概括
这项研究发现,带有压力映射传感器的闭合力测量装置对老年人来说是非常可靠的. 该设备提供一致的测量,无论由同一检查员多次或由不同的检查员进行.
科学领域:
- 老年学是一门学科.
- 生物医学工程 生物医学工程
- 牙科科学 牙科科学
背景情况:
- 评估闭塞力对于评估老年人口腔健康至关重要.
- 测量工具的可靠性对于准确的临床评估和研究至关重要.
- 之前的设备可能在准确性或易于使用方面存在局限性.
研究的目的:
- 为了确定特定的闭塞力测量装置的内部检查员和内部检查员的可靠性.
- 为了评估电容型压力测绘传感器表在测量老年人的闭塞力时的性能.
主要方法:
- 共有38名65岁及以上的参与者被录取.
- 使用Oramo-bf装置测量了闭塞力.
- 内部检查员的可靠性由一名牙医评估,每位参与者进行三次测量,计算类内相关系数 (ICC) 和变化系数 (CV).
- 三名牙医评估了检查者之间的可靠性,每位参与者测量一次闭塞力,计算ICC.
主要成果:
- 该设备显示出高的内部检查员可靠性,单次测量时的ICC值为0.982,三次测量的平均值为0.994.
- 平均变化系数 (CV) 是0.052±0.046,表明变化很低.
- 观察到高的相互检查员可靠性,ICC为0.950.
结论:
- 带有电容型压力映射传感器的闭塞力测量装置在老年人中表现出极好的内部和内部检查员可靠性.
- 这些发现表明,这种设备可以产生可靠的闭塞力测量,即使只读取一次.
- 这种工具有望在老年牙科中提供可靠的临床和研究应用.
相关概念视频
Pressure Gauges
5.4K
Most pressure gauges, like those on scuba tanks, are calibrated to read zero at atmospheric pressure. Readings from such gauges are called the gauge pressure, which is the pressure relative to atmospheric pressure. When the pressure inside the tank exceeds atmospheric pressure, the gauge reports a positive value. Some gauges are designed to measure negative pressure. For example, many physics experiments must take place in a vacuum chamber, a rigid chamber from which some of the air is pumped...
5.4K
Measurement of Fluid Pressure
2.3K
Fluid pressure is commonly measured using devices called manometers, which rely on liquid columns to indicate pressure differences. The height of a liquid column in a manometer reflects the pressure exerted by the fluid, providing a simple yet effective means of measurement. Different types of manometers serve specific purposes based on their configurations and the type of fluids involved.
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
2.3K
Bioreactor Controls-I
103
Maintaining optimal conditions within fermenters is essential for maximizing microbial productivity and ensuring process efficiency. This lesson focuses on key parameters—temperature, foam, pH, carbon dioxide, oxygen, and pressure—and their precise measurement and control strategies in fermentation systems.Temperature ControlTemperature regulation is critical due to the exothermic nature of many fermentation processes. In small laboratory fermenters, temperature is commonly...
103