临床定量方法 开发使用LC-MS/MS的血清氧硫酸盐 (IS) 测定方法
Sohei Kobayashi1,2, Kazuyuki Matsushita2, Kenji Kawasaki2
1Department of Medical Technology & Sciences, School of Health Sciences at Narita, International University of Health and Welfare, Chiba, Japan.
Biochemistry and biophysics reports
|November 10, 2025
概括
硫酸 (IS) 是功能下降的标志物. 这项研究验证了液体染色学-并列质谱法 (LC-MS/MS) 方法用于标准化IS测定,显示出临床使用的优异性能和可重复性.
科学领域:
- 临床化学 临床化学
- 分析化学 分析化学
- 生物标记分析 生物标记分析
背景情况:
- 硫酸 (IS) 是一种尿素毒素,随着功能下降而积累,加剧病的进展.
- 目前的IS测试试剂由于缺乏标准测试方法而缺乏标准化.
- 建立可靠的IS测定对于监测功能和疾病管理至关重要.
研究的目的:
- 评估液体染色学-并联质谱法 (LC-MS/MS) 的性能,以建立氧硫酸盐 (IS) 的标准检测方法.
- 评估该方法作为临床环境中IS量化参考标准的适用性.
主要方法:
- 使用AB SCIEX QTRAP® 4500系统进行LC-MS/MS分析.
- 方法验证包括线性,可重现性,选择性,干扰测试和LOB,LOD和LLOQ的确定.
- 通过使用IS的峰值面积与其化内部标准 (IS-d4) 的比率生成一个校准曲线.
主要成果:
- 通过LC-MS/MS方法证明了极好的线性 (R2=0.9995) 到440μmol/L.
- 观察到高精度的变化系数为2.6-4.7%的可重复性和7.9-9.2%的中间精度.
- 该方法显示了高回收 (101.0-104.3%),最小干扰 (<5.0%),并确定了LOB (0.004μmol/L),LOD (1.248μmol/L) 和LOQ (3.23μmol/L).
- 在LC-MS/MS方法和常规酶分析之间发现了强烈的相关性 (R = 0.996).
结论:
- 经过验证的LC-MS/MS方法表现出令人满意的基本性能和出色的精度,表明了试验稳定性.
- 这种方法适用于建立标准化的IS测定.
- 这些发现支持其作为临床诊断中氧硫酸量化可靠参考方法的实用性.
更多相关视频
08:53Quantitative and Qualitative Method for Sphingomyelin by LC-MS Using Two Stable Isotopically Labeled Sphingomyelin Species
Published on: May 7, 2018
10.2K
08:06Absolute Quantification of Cell-Free Protein Synthesis Metabolism by Reversed-Phase Liquid Chromatography-Mass Spectrometry
Published on: October 25, 2019
9.6K
相关概念视频
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then passed on to...
Atomic Emission Spectroscopy: Lab
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...