Jove
Visualize
联系我们
JoVE
x logofacebook logolinkedin logoyoutube logo
关于 JoVE
概览领导团队博客JoVE 帮助中心
作者
出版流程编辑委员会范围与政策同行评审常见问题投稿
图书馆员
用户评价订阅访问资源图书馆顾问委员会常见问题
研究
JoVE JournalMethods CollectionsJoVE Encyclopedia of Experiments存档
教育
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab Manual教师资源中心教师网站
使用条款与条件
隐私政策
政策

相关概念视频

Pulse rhythm01:30

Pulse rhythm

1.3K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.3K
Holter Monitor: 24-Hour Monitoring01:23

Holter Monitor: 24-Hour Monitoring

1.9K
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
1.9K
Errors occurring during blood pressure monitoring01:25

Errors occurring during blood pressure monitoring

1.3K
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
1.3K
Cardiomyopathy V: Interprofessional Care01:29

Cardiomyopathy V: Interprofessional Care

323
Managing cardiomyopathy involves addressing underlying or precipitating causes, treating heart failure with medications, and implementing dietary changes and a balanced exercise and rest regimen.Lifestyle ModificationsCardiomyopathy patients should adopt a low-sodium diet to reduce fluid retention and manage heart failure. A personalized exercise and rest plan helps maintain physical fitness without overstraining the heart. Avoiding alcohol and tobacco is essential to prevent further damage to...
323

您也可能阅读

相关文章

通过共同作者、期刊和引用图与本文相关的文章。

排序
Same author

Artificial intelligence vs electrophysiologist adjudication of atrial arrhythmias.

Heart rhythm·2026
Same author

Safety, performance, and remote programming utilization of an insertable cardiac monitor: The LUX-Dx PERFORM study.

Heart rhythm O2·2026
Same author

Spatiotemporal mapping for detection of anatomic locations of drivers of atrial fibrillation in swine.

Heart rhythm O2·2026
Same author

Frequent Droughts Reduce Carbon Stabilisation in Organo-Mineral Soils.

Global change biology·2026
Same author

Premature ventricular contraction detection and estimation of daily burden by an insertable cardiac monitor.

Heart rhythm O2·2025
Same author

KRIT1 heterozygous mutations are sufficient to induce a pathological phenotype in patient-derived iPSC models of cerebral cavernous malformation.

Cell reports·2025
Same journal

Dual Coronary Sinus Lead Strategy to Avoid Tricuspid Valve Traversal in Biventricular Pacing.

Pacing and clinical electrophysiology : PACE·2026
Same journal

A Case of Permanent Pacemaker Implantation via the Epicardial Approach Using the 3830 Lead in an 11-Day-Old Neonate (With Follow-Up of the Above Case).

Pacing and clinical electrophysiology : PACE·2026
Same journal

Cryoballoon Versus Radiofrequency Ablation for Persistent Atrial Fibrillation: Meta-Analysis of Randomized Trials.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Tilt Test Duration in Suspected Vasovagal Syncope: Temporal Patterns and Diagnostic Yield in Patients From Central China.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Combined Leadless Pacing and Subcutaneous ICD Therapy in Long QT Syndromes.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Association of Anesthesia Modality With Procedural Parameters and Clinical Outcomes in PVI for Atrial Fibrillation.

Pacing and clinical electrophysiology : PACE·2026
查看所有相关文章

相关实验视频

Updated: Jan 11, 2026

A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
04:24

A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program

Published on: April 19, 2019

12.2K

插入式心脏监测器的算法改进减少了假阳性和节目审查负担.

Mark Richards1, Molly Kupfer2, Keith Herrmann2

  • 1Baylor Scott & White Arrhythmia Management, Plano, Texas, USA.

Pacing and clinical electrophysiology : PACE
|November 10, 2025
PubMed
概括
此摘要是机器生成的。

新的可插入心脏监测 (ICM) 算法显著减少了对心房动 (AF) 和暂停的错误阳性. 优化设置还可以减少非可操作的夜间肌,改善设备性能和诊所工作流程.

关键词:
节律失常 (arrhythmia) 是一种心律失常.心房动是心房动的一种.胸肌梗塞是什么意思 胸肌梗塞是什么意思插入式心脏监测器是可以插入的心脏监测器.暂停暂停暂停暂停暂停

更多相关视频

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
05:03

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function

Published on: December 11, 2019

9.0K
A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
18:11

A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis

Published on: December 28, 2012

24.7K

相关实验视频

Last Updated: Jan 11, 2026

A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
04:24

A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program

Published on: April 19, 2019

12.2K
Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
05:03

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function

Published on: December 11, 2019

9.0K
A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
18:11

A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis

Published on: December 28, 2012

24.7K

科学领域:

  • 心脏病学 心脏病学
  • 生物医学工程 生物医学工程
  • 医疗器械 医疗器械

背景情况:

  • 插入式心脏监测器 (ICM) 越来越多地用于检测心律失常.
  • 目前的ICM面临的挑战是性能准确性和数据管理.
  • LUX-Dx II+ ICM 结合了对心房动 (AF),暂停和胸检测的改进.

研究的目的:

  • 评估LUX-Dx II+ ICM中增强的AF和暂停检测算法的影响.
  • 评估优化夜间编程的有效性,以减少非可操作的胸肌梗塞和暂停检测.

主要方法:

  • 现实世界的心电图数据被用来比较新的AF和暂停算法与前身LUX-Dx算法.
  • 模拟优化的夜间编程设置 (晚上11点至早上7点) 应用于LUX-Dx数据.
  • 评估了非可采取行动的夜间胸 (>30bpm) 和暂停 (<5秒) 的减少.

主要成果:

  • 新的AF算法减少了34-38%的假阳性 (在敏感设置中高达74%) ,同时保持了>98%的灵敏度.
  • 增强的暂停算法将假阳性减少了48.6-62.5%,相对灵敏度为100%.
  • 调整后的夜间值减少了心和停顿发作分别为98%和90% (总体为75%和56%).

结论:

  • 改进的ICM算法和编程设置显著减少了假阳性AF和暂停检测.
  • 优化的设置大大降低了非可操作的夜间心肌梗塞和暂停检测.
  • 这些改进有望提高ICM的性能,并简化设备诊所的工作流程.