1Organic Metrology Group, Division of Chemical and Material Metrology, Korea Research Institute of Standards and Science, 267 Gajeong-ro, Yuseong-gu, Daejeon 34113, Republic of Korea; Department of Precision Measurement, University of Science and Technology, 217 Gajeong-ro, Yuseong-gu, Daejeon 34113, Republic of Korea.
Journal of chromatography. A
|November 21, 2025
概括
本研究提出了一种新方法,用于精确测量各种样本中的14种氨基酸 (AAs),使用液态染色学-并联质谱法 (LC-MS/MS) 和同位素稀释质谱法 (IDMS). 经过验证的技术确保了在各种应用中准确的AA量化.
科学领域:
- 分析化学 分析化学
- 生物化学 生物化学
- 质谱测量质量谱测量
背景情况:
- 精确量化氨基酸 (AAs) 对于各种生物和临床应用至关重要.
- 现有的方法可能缺乏对不同矩阵所需的精度和范围.
- 对于未衍生型AAs而言,需要一个可靠的参考测量程序.
研究的目的:
- 开发和验证一种新的参考测量程序,用于同时量化14种未衍生氨基酸.
- 建立一种可靠的方法,适用于各种矩阵,包括等离子体等生物样本.
- 为了确保氨基酸测量的高准确性和精度.
主要方法:
- 离子配对液体色谱-并联质谱法 (LC-MS/MS) 与同位素稀释质谱法 (IDMS) 结合.
- 在6分钟内使用三酸和七酸在逆相柱上分离未衍生AA.
- 采用精确匹配的双IDMS与同位素标记的AA作为内部标准.
主要成果:
- 在缓冲和等离子矩阵中成功分离和定量14种未衍生氨基酸.
- 该方法在1至20毫克/公斤度范围内显示出高准确度和精度.
- 验证证实了线性,灵敏度,准确性和精度,其可靠性由经过认证的参考材料证实.
结论:
- 开发的LC-MS/MS-IDMS方法适用于在各种矩阵中准确量化14种氨基酸.
- 这种程序可以作为氨基酸测量可靠的初级方法.
- 该方法的稳定性支持其在认证参考材料和其他分析需求中的应用.
相关概念视频
Peptide Identification Using Tandem Mass Spectrometry
8.1K
Tandem mass spectrometry, also known as MS/MS or MS2, is an analytical technique that employs two mass analyzers. Essentially it is a series of mass spectrometers that helps isolate a particular biomolecule and then helps study its chemical properties.
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
8.1K
Tandem Mass Spectrometry
2.3K
Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and reduce chemical noise during analyte detection. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called...
2.3K
Mass Spectrometry of Amines
5.3K
In mass spectroscopy, amines undergo fragmentation to give parent ions with odd molecule weights. This observed mass spectrum follows the nitrogen rule; a molecule with an odd number of nitrogen atoms produces a molecular ion with an odd molecular weight. Amines undergo fragmentation through α cleavage, producing nitrogen-containing cations—iminium ions—and alkyl radicals. Mass spectra of aromatic and cyclic aliphatic amines exhibit strong molecular ion peaks, but acyclic...
5.3K