自动备用光源技术可以显著降低在内镜手术过程中患者受伤和外科窗烧伤的风险
Matthew J Kabalan1, Tyler Merrill1, Raj Sindwani1
1Department of Otolaryngology-Head & Neck Surgery, Section of Rhinology, Sinus & Skull Base Surgery, Cleveland Clinic, Cleveland, OH, USA.
American journal of rhinology & allergy
|November 25, 2025
概括
自动备用技术 (AST) 在内镜手术期间显著降低烧伤风险. 这种安全功能快速冷却断开的光线电缆,防止患者受伤和手术窗发生火灾.
科学领域:
- 医疗器械 医疗器械
- 外科手术的安全性
- 内镜技术的使用方法
背景情况:
- 烧伤和手术窗火灾是内镜手术中的风险.
- 自动待机技术 (AST) 旨在通过在断开时禁用光源来减轻这些风险.
研究的目的:
- 为了比较AST与传统光源的安全性.
- 为了评估内镜尖端和适配器的温度差异.
- 为了评估被断开的光线电缆点燃外科手术窗的倾向.
主要方法:
- AST和常规光源被连接到刚性鼻腔内镜.
- 用热电偶测量内镜尖端和适配器的温度.
- 一个热摄像头评估了光线电缆在断开后的温度.
- 断开的光线电缆被测试了它们燃烧手术窗的能力.
主要成果:
- AST导致内镜尖端 (24.0°C与25.2°C) 和适配器 (26.6°C与27.3°C) 的平均温度较低.
- 与AST (40.9°C) 相比,断开的传统光线电缆达到更高的温度 (56.8°C).
- 传统电缆造成了显著的窗烧伤,而AST连接的电缆没有.
结论:
- AST技术在内镜手术中提供了相当大的安全益处.
- 它有效地减少了患者和护理人员因热暴露而受到伤害的风险.
- 通过快速冷却断开的光线电缆,AST可以防止外科手术窗发生火灾.
相关概念视频
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
526
Sigmoidoscopy and laparoscopy are distinct medical procedures that enable physicians to internally inspect different parts of the GI tract. Although they serve different purposes, each is essential for diagnosing and, in some cases, treating various medical conditions.
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
526
Endoscopic Procedures I: Esophagogastroduodenoscopy
841
An Esophagogastroduodenoscopy (EGD) is a diagnostic procedure in which an endoscopist uses a flexible, lighted endoscope to visualize the upper gastrointestinal (GI) tract. The procedure includes visualizing the oropharynx, esophagus, stomach, and the first part of the small intestine, the duodenum.
During an EGD, the endoscope can be used to:
During an EGD, the endoscope can be used to:
841