在表面灭菌中测量蒸汽成分的标准方法的定量评估
Francesco Tessarolo1,2, Michela Masè3
1Department of Industrial Engineering, University of Trento, via delle Regole, 101, Trento, 38123, Italy. francesco.tessarolo@unitn.it.
Scientific reports
|December 1, 2025
概括
在蒸汽灭菌中测量非可冷凝气体 (NCG) 的标准方法由于未指定的参数而不可靠. 优化测试程序对于准确的蒸汽成分评估和杀菌器验证至关重要.
科学领域:
- 医疗器械 医疗器械
- 消毒技术 消毒技术 消毒技术
- 质量控制 质量控制
背景情况:
- 蒸汽成分,特别是非可冷凝气体 (NCG) 的部分,对于有效的蒸汽杀菌至关重要.
- 标准EN 285:2015+A1:2021提供了测量NCG分数 (NCG%) 的方法,但缺乏工艺参数和环境条件的详细规格.
研究的目的:
- 量化未指定的参数对NCG%测量的影响.
- 在蒸汽杀菌试验中确定影响NCG%变量的因素.
- 为更可靠的蒸汽杀菌试验程序提出改进建议.
主要方法:
- 具有不同配置和负载的灭菌周期是在中型蒸汽灭菌器中进行的.
- 非可冷凝气体和蒸汽冷凝液体积在30秒间隔间进行测量.
- 在不同的时间点和凝结体量计算NCG%值,并对环境条件影响进行理论估计.
主要成果:
- NCG%值和测试结果受到工艺配置文件,负载量和凝结物收集时间/体积的显著影响.
- 环境条件 (温度和压力) 显示对NCG%测量产生不可忽视的影响.
- 在NCG测量的变化损害了杀菌器资格和蒸汽组成估计的标准方法的可靠性.
结论:
- 标准方法中缺乏详细规范导致NCG%的测量变化,影响消毒器的资格.
- 建议在生产过程中实时对蒸汽成分进行定量监测.
- 改进的测试程序是必要的,以提高蒸汽杀菌的有效性和质量保证.
相关概念视频
Methods of Sterilization I: Physical Methods
23.3K
As used in a healthcare facility, sterilization destroys all microorganisms through physical or chemical methods. The physical method includes steam, dry heat, boiling water, and radiation.
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
23.3K
Methods of Sterilization II: Chemical Methods
8.8K
In healthcare, the chemical method of sterilization uses chemical sterilants to treat surgical instruments and medical supplies to help prevent the transmission of infectious pathogens to patients. Due to heat sensitivity, most medical supplies and equipment should not be exposed to high temperatures. These parts include rubber, plastic, glass, and other similar elements.
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
8.8K
Standard Solutions
2.3K
Standard solutions refer to solutions with a precisely known concentration or composition. A primary standard is a highly pure, high molar mass, stable substance that is entirely soluble in water, the most commonly used solvent in analytical chemistry. The primary standard solution can be used to standardize secondary standards, which are substances with known concentrations but are less pure and stable. Standard solutions are essential for achieving accurate and reliable results in analytical...
2.3K
Atomic Absorption Spectroscopy: Lab
979
For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
979
Quantitative Analysis
1.2K
Quantitative analysis is a technique for measuring the amount of specific constituents in a sample. When the sample's composition is unknown, qualitative analysis is performed first to identify its components, which ensures that the correct substances are measured during the quantitative phase.
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
1.2K
Measurement: Standard Units
77.4K
Every measurement provides three kinds of information: the size or magnitude of the measurement (a number), a standard of comparison for the measurement (a unit), and an indication of the uncertainty of the measurement. While the number and unit are explicitly represented when a quantity is written, the uncertainty is an aspect of the errors in the measurement results.
77.4K