OPSensing:采用口腔内可穿戴设备实现实时PPG监控
概括
介绍OPSensing,一个创新的口内可穿戴设备,用于持续的生理监测. 该系统提供精确的脉冲率和血液氧和度测量,克服了传统可穿戴设备的局限性.
科学领域:
- 生物医学工程 生物医学工程
- 可穿戴技术可穿戴技术
- 生理监测 生理监测
背景情况:
- 摄影电磁镜 (PPG) 对于监测脉冲率 (PR) 和血液氧和 (SpO2) 至关重要.
- 现有的可穿戴设备面临着运动工件,噪音和用户舒适性方面的挑战.
- 口腔环境为PPG信号采集提供了一个稳定,屏蔽的位置.
研究的目的:
- 开发和评估OPSensing,一个口内PPG传感系统.
- 为了实现实时,准确的PR和SpO2估计.
- 为了证明口内可穿戴设备在持续生理监测中的潜力.
主要方法:
- 开发了一个口内PPG传感系统 (OPSensing).
- 使用多波长PPG方法和信号处理算法.
- 收集了10名参与者的数据,与商用脉冲氧计进行了比较.
主要成果:
- OPSensing实现了PR的根平均平方误差 (RMSE) 为1.88 bpm,而SpO2.2的误差为2.99%.
- 结果符合美国食品和药物管理局 (FDA) 推的RMSE值.
- 在睡眠研究中,在七个小时内表现出连续监控能力.
结论:
- OPSensing为准确,非侵入性生理监测提供了一个可行的解决方案.
- 口内环境增强了PPG信号质量,减少了人工物.
- 内口可穿戴设备显示出长期健康跟踪的巨大潜力.
相关概念视频
Pulse rhythm
1.3K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.3K
Holter Monitor: 24-Hour Monitoring
1.9K
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
1.9K
Pulse Oximetry
1.2K
Pulse oximetry, or SpO2, is a non-invasive method for continuously monitoring arterial oxygen saturation (SaO2). This procedure involves attaching a probe or sensor to the patient's fingertip, forehead, earlobe, or nose bridge. The sensor works by detecting changes in oxygen saturation levels through light signals generated by the oximeter and reflected by the pulsing blood under the probe.
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
1.2K
Equipments Used To Measure Blood Pressure
3.0K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.0K