使用前段光学连贯断层扫描和Scheimpflug成像,在角膜眼中加速角膜交叉连接后,对角膜平衡演变的比较
Antonia Lang1, Berthold Seitz2, Cristian Munteanu2
1Department of Ophthalmology, Saarland University Medical Center, Kirrberger Straße, Building 22, 66424, Homburg, Germany. anla00014@stud.unisaarland.de.
International ophthalmology
|December 5, 2025
概括
Pentacam HR Scheimpflug成像检测到角膜在加速角膜交叉连接 (A-CXL) 后角膜稀薄,但Casia 2 AS-OCT没有. 这些测量设备对于监测A-CXL结果是不可互换的.
科学领域:
- 眼科医生 眼科 眼科
- 角质科学 角质科学
- 医疗成像医学成像
背景情况:
- 角膜 (KC) 是角膜逐渐变薄的疾病.
- 加快角膜交叉连接 (A-CXL) 是一种治疗方法,用于硬角膜并阻止KC的进展.
- 精确的精度测量对于监测治疗疗效至关重要.
研究的目的:
- 为了比较A-CXL后使用Pentacam HR (Scheimpflug) 和Casia 2 (AS-OCT) 的纵向角膜平衡测量.
- 评估这两种成像方法在测量A-CXL后的中央 (CCT) 和最薄 (TCT) 角膜厚度时的一致性.
主要方法:
- 对218个用A-CXL治疗的KC眼睛进行了回顾性研究.
- 用Pentacam HR和Casia 2在多个时间点测量角膜厚度,直到最近的随访.
- 在A-CXL期间用于监测CCT的手术内超声波平衡.
主要成果:
- 在A-CXL后,Pentacam HR显示出显著的CCT和TCT减少,而Casia 2则没有.
- 在A-CXL后六周观察到设备之间的最高测量一致性.
- 角膜密度测量在六周达到峰值,然后下降,但没有达到A-CXL前的水平.
结论:
- 在A-CXL之后,Pentacam HR和Casia 2的精度测量是不可互换的.
- 在A-CXL之后,基于Scheimpflug的厚度测量需要谨慎的解释,特别是在第一年.
- 由Pentacam HR观察到的角膜变薄,Casia 2没有检测到,突出了设备特异性的差异.
相关概念视频
Phase Contrast and Differential Interference Contrast Microscopy
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Confocal Fluorescence Microscopy
Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...